The Role of a Novel Digital Microcapillary Instrument in Detecting Blood and Plasma Hyperviscosity

ACTA MEDICA INDONESIANA · Vol. 46 No. 3 (2016)

ISSN
2338-2732, 0125-9326
Terbit
2016-05-16
Penerbit
The Indonesian Society of Internal Medicine
Bahasa
eng

Abstrak

Aim: to test the precision and accuracy of a Digital Microcapillary instrument in measuring blood and plasma viscosity. Methods: about 40 blood samples were drawn for precision test. The samples were obtained from patients admitted for Medical Check-Up at CiptoMangunkusumo Hospital (RSCM) in December 2011. Accuracy test was evaluated using cross-sectional design and involving 135 patients with acute ischemic stroke. The patients underwent either inpatients or outpatients care at Department of Clinical Pathology, Department of Neurology,and Emergency Unit of RSCM, Fatmawati Hospital Jakarta, Prikasih Hospital Jakarta, and Bhakti Yuda Hospital Depok. The precision test was evaluated by calculating the coefficient of variation (CV), interrater variability of Cronbach Alpha, and reliability coefficient of Bland Altman. The accuracy of the test was evaluated with a diagnostic test. The gold standard used for these tests was Brookfield LV-DV III viscometer. Results: the results of precision test were: CV = 0.04; interrater variability of blood and plasma viscosity = 0.94 and 0.82, respectively; the Bland Altman mean difference = –0.19. The results of accuracy test were: sensitivity of blood and plasma viscosity measurement were 88.9% and 100%, respectively; specificity of blood and plasma viscosity measurement were 88.9% and 84%, respectively. Conclusion: the digital microcapillary has high sensitivity and specificity; therefore the instrument can be considered to be used as screening test tool to measure blood and plasma viscosity. Key words: digital microcapillary, blood viscosity, plasma viscosity, acute ischemic stroke

Jaringan kolaborasi para penulis

40 penulis · 146 tautan kepenulisan bersama · menampilkan 40 yang paling banyak terhubung

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Sebelas Maret University
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Andhika RachmanAfiliasi tidak diketahui3
Cosphiadi IrawanAfiliasi tidak diketahui3
Karmel L TambunanAfiliasi tidak diketahui2
Ikhwan RinaldiAfiliasi tidak diketahui2
Rahmat CahyanurAfiliasi tidak diketahui2
Achmad Fauzi KamalAfiliasi tidak diketahui2
Marlinda AdhamAfiliasi tidak diketahui2
Mardiah Suci HardiantiAfiliasi tidak diketahui2
Thariqah SalamahAfiliasi tidak diketahui2
Evy YunihastutiAfiliasi tidak diketahui2
Simon SalimAfiliasi tidak diketahui2
Sally Aman NasutionAfiliasi tidak diketahui2
Nurul Inayah RahmaniAfiliasi tidak diketahui2
Judhiastuty FebruhartantyAfiliasi tidak diketahui1
Fitya Safira BirahmatikaAfiliasi tidak diketahui1
Khalida FauziaAfiliasi tidak diketahui1
Indriya Laras PramesthiAfiliasi tidak diketahui1
Helda KhusunAfiliasi tidak diketahui1
Hera NurlitaAfiliasi tidak diketahui1
Ahmad Thohir HidayatAfiliasi tidak diketahui1
Teguh BudiharjoAfiliasi tidak diketahui1
Djoko PriyatnoAfiliasi tidak diketahui1
Anung SugihantonoAfiliasi tidak diketahui1
Yusuf PerdanaAfiliasi tidak diketahui1
Rinaldo Adi PratamaAfiliasi tidak diketahui1
Sumargono SumargonoAfiliasi tidak diketahui1
Sukman Tulus PutraAfiliasi tidak diketahui1
Sudigdo SatroasmoroAfiliasi tidak diketahui1
Andri Maruli T LubisAfiliasi tidak diketahui1
Kurniadi HusodoAfiliasi tidak diketahui1
Djajadiman GatotAfiliasi tidak diketahui1
T. Djumhana AtmakusumaAfiliasi tidak diketahui1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) http://www.actamedindones.org/index.php/ijim/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 8 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">The Role of a Novel Digital Microcapillary Instrument in Detecting Blood and Plasma Hyperviscosity</dc:title>
	<dc:creator>Rasyid, Al</dc:creator>
	<dc:creator>Misbach, Jusuf</dc:creator>
	<dc:creator>Purba, Jan S</dc:creator>
	<dc:creator>Timan, Ina S</dc:creator>
	<dc:creator>Sukrisman, Lugyanti</dc:creator>
	<dc:creator>Mansyur, Muchtaruddin</dc:creator>
	<dc:creator>Yudiarsah, Efta</dc:creator>
	<dc:creator>Suroto, Suroto</dc:creator>
	<dc:description xml:lang="en-US">Aim: to test the precision and accuracy of a Digital Microcapillary instrument in measuring blood and plasma viscosity. Methods: about 40 blood samples were drawn for precision test. The samples were obtained from patients admitted for Medical Check-Up at CiptoMangunkusumo Hospital (RSCM) in December 2011. Accuracy test was evaluated using cross-sectional design and involving 135 patients with acute ischemic stroke. The patients underwent either inpatients or outpatients care at Department of Clinical Pathology, Department of Neurology,and Emergency Unit of RSCM, Fatmawati Hospital Jakarta, Prikasih Hospital Jakarta, and Bhakti Yuda Hospital Depok. The precision test was evaluated by calculating the coefficient of variation (CV), interrater variability of Cronbach Alpha, and reliability coefficient of Bland Altman. The accuracy of the test was evaluated with a diagnostic test. The gold standard used for these tests was Brookfield LV-DV III viscometer. Results: the results of precision test were: CV = 0.04; interrater variability of blood and plasma viscosity = 0.94 and 0.82, respectively; the Bland Altman mean difference = –0.19. The results of accuracy test were: sensitivity of blood and plasma viscosity measurement were 88.9% and 100%, respectively; specificity of blood and plasma viscosity measurement were 88.9% and 84%, respectively. Conclusion: the digital microcapillary has high sensitivity and specificity; therefore the instrument can be considered to be used as screening test tool to measure blood and plasma viscosity. Key words: digital microcapillary, blood viscosity, plasma viscosity, acute ischemic stroke</dc:description>
	<dc:publisher xml:lang="en-US">The Indonesian Society of Internal Medicine</dc:publisher>
	<dc:date>2016-05-16</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>http://www.actamedindones.org/index.php/ijim/article/view/92</dc:identifier>
	<dc:source xml:lang="en-US">Acta Medica Indonesiana; Vol. 46 No. 3 (2014)</dc:source>
	<dc:source>2338-2732</dc:source>
	<dc:source>0125-9326</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>http://www.actamedindones.org/index.php/ijim/article/view/92/88</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2016 Indonesian Journal of internal Medicine</dc:rights>
</oai_dc:dc>