Determination of the Diagnostic Reference Level (DRL) in Samarinda Hospitals

Atom Indonesia · Vol. 49 No. 3 (2023) · hlm. 145–150

DOI
10.55981/aij.2023.1285
ISSN
2356-5322, 0126-1568
Terbit
2023-11-23
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

The diagnostic reference level (DRL) is a form of investigative level used as a tool to help optimize protection to radiation exposure for diagnostic and interventional procedures. The purpose of this study was to determine the local DRL values for the examination of the abdomen, thorax, and head at radiology installations. The modality used was 128-slice CT scan. The numbers of patients whose data were used were 200 for abdominal examinations, 160 for thoracic examinations, and 100 for head examinations. Overall, the total patient whose data was used was 460. Data processing in this study was carried out with a quantitative analysis technique, namely descriptive statistics. This analysis technique used secondary data obtained from the results of recaptures or archival books for examination of the abdomen, thorax, and head. Data processing was carried out with a measure of diversity through the calculation of the third quartile in the data distribution. It was assumed that 75 % of patients performed examinations with a common diagnosis. The results of these calculations are visualized in the form of graphs of the relationship of computed tomography dose index volume (CTDIvol) with the number of patients and a graph of the relationship of dose length product (DLP) with the number of patients. In the abdominal examination, a CTDIvol of 12 mGy and a DLP of 1545.5 mGy·cm. In the thoracic examination, a CTDIvol of 11 mGy and a DLP of 903 mGy·cm were obtained. For the head examination, a CTDIvol of 34.25mGy and a DLP of 2190.25 mGy·cm were obtained. The conclusion obtained from this study is that the DRLs are relatively low, but they still need to be optimized by medical physicists.

Kata kunci

Jaringan kolaborasi para penulis

23 penulis · 46 tautan kepenulisan bersama · 6 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Mulawarman University
  • Lambung Mangkurat University
  • University of Indonesia
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Sahara Hamas IntifadhahMulawarman University4
A. PutriAfiliasi tidak diketahui2
Retno ZurmaAfiliasi tidak diketahui2
Amar Vijai NasrullohLambung Mangkurat University1
Arfan Eko FahrudinLambung Mangkurat University1
M. HamidMulawarman University1
A. Mu'tiAfiliasi tidak diketahui1
A. NurMulawarman University1
Kholis NurhanafiMulawarman University1
Sherly PutriMulawarman University1
Fatimah Kunti HentihuAfiliasi tidak diketahui1
D. R. (Dian) PutriMulawarman University1
Robert Janssen StevenlyAfiliasi tidak diketahui1
R. AnthonMulawarman University1
Prawito PrajitnoUniversity of Indonesia1
Djarwani Soeharso SoejokoUniversity of Indonesia1
Ahmad ZarkasiMulawarman University1
Dadan HamdaniMulawarman University1
Adrianus Inu NatalisantoMulawarman University1
Suhadi MuliyonoMulawarman University1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Determination of the Diagnostic Reference Level (DRL) in Samarinda Hospitals</dc:title>
	<dc:creator>Jannah, R.</dc:creator>
	<dc:creator>Munir, R.</dc:creator>
	<dc:creator>Putri, E. R.</dc:creator>
	<dc:subject xml:lang="en-US">Abdomen; CT-Scan; Diagnostic Reference Level (DRL); Head; Thorax</dc:subject>
	<dc:description xml:lang="en-US">The diagnostic reference level (DRL) is a form of investigative level used as a tool to help optimize protection to radiation exposure for diagnostic and interventional procedures. The purpose of this study was to determine the local DRL values for the examination of the abdomen, thorax, and head at radiology installations. The modality used was 128-slice CT scan. The numbers of patients whose data were used were 200 for abdominal examinations, 160 for thoracic examinations, and 100 for head examinations. Overall, the total patient whose data was used was 460. Data processing in this study was carried out with a quantitative analysis technique, namely descriptive statistics. This analysis technique used secondary data obtained from the results of recaptures or archival books for examination of the abdomen, thorax, and head. Data processing was carried out with a measure of diversity through the calculation of the third quartile in the data distribution. It was assumed that 75 % of patients performed examinations with a common diagnosis. The results of these calculations are visualized in the form of graphs of the relationship of computed tomography dose index volume (CTDIvol) with the number of patients and a graph of the relationship of dose length product (DLP) with the number of patients. In the abdominal examination, a CTDIvol of 12 mGy and a DLP of 1545.5 mGy·cm. In the thoracic examination, a CTDIvol of 11 mGy and a DLP of 903 mGy·cm were obtained. For the head examination, a CTDIvol of 34.25mGy and a DLP of 2190.25 mGy·cm were obtained. The conclusion obtained from this study is that the DRLs are relatively low, but they still need to be optimized by medical physicists.</dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US"></dc:contributor>
	<dc:date>2023-11-23</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1285</dc:identifier>
	<dc:identifier>10.55981/aij.2023.1285</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; VOL 49, NO 3 (2023): DECEMBER 2023; 145-150</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v49i3</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1285/1092</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1285/2349</dc:relation>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>