Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates

Atom Indonesia · Vol. 49 No. 1 (2023) · hlm. 53–59

DOI
10.55981/aij.2023.1261
ISSN
2356-5322, 0126-1568
Terbit
2023-04-14
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

Continuous gantry motion, continuous beam modulation, and variable dose rate are used in volumetric modulated arc therapy (VMAT) to obtain highly conformal radiation therapy dose distributions. Several errors during daily radiation therapy treatment can be sources of uncertainties in dose delivery. These errors include monitor unit calculation errors and other human mistakes. Due to the uncertainties in the excessively modulated VMAT plan, the intended dose distribution is not delivered perfectly, leading to a mismatch between the measured and planned dose distributions. This necessitates an extensive and effective quality assurance (QA) program for both machine and patient. In this study, VMAT QA plan verification of 62 head and neck (HN) and 19 prostate cases was done using Octavius 4D setup with its associating VeriSoft gamma analysis software. The plans showed a maximum 3D gamma passing rate with 4 mm/3 % gamma acceptance criteria, i.e., 99.7 % for the HN cancer cases and 99.5 % for the prostate cancer cases. Local gamma analysis was also performed for both regions. Furthermore, 2D and volumetric gamma analyses were also carried out. Gamma analysis with respect to different axis was also carried out. It was known that the transversal axis showed the highest gamma passing rate in both HN and prostate cases, i.e., 99.17 % and 98.3 %, respectively. The transverse axis came to be a better fit for the planned dose distribution.

Kata kunci

Jaringan kolaborasi para penulis

4 penulis · 6 tautan kepenulisan bersama

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Pakistan Institute of Engineering and Applied Sciences
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Patient-Specific Pre-Treatment VMAT Plan Verification Using Gamma Passing Rates</dc:title>
	<dc:creator>Arsalan, M. Z.</dc:creator>
	<dc:creator>Kakakhel, M. B.</dc:creator>
	<dc:creator>Shamshad, M.</dc:creator>
	<dc:creator>Afridi, T. A.</dc:creator>
	<dc:subject xml:lang="en-US">VMAT, Treatment Planning System Gamma Analysis Octavius 4D Setup Gamma Acceptance Criteria</dc:subject>
	<dc:description xml:lang="en-US">Continuous gantry motion, continuous beam modulation, and variable dose rate are used in volumetric modulated arc therapy (VMAT) to obtain highly conformal radiation therapy dose distributions. Several errors during daily radiation therapy treatment can be sources of uncertainties in dose delivery. These errors include monitor unit calculation errors and other human mistakes. Due to the uncertainties in the excessively modulated VMAT plan, the intended dose distribution is not delivered perfectly, leading to a mismatch between the measured and planned dose distributions. This necessitates an extensive and effective quality assurance (QA) program for both machine and patient. In this study, VMAT QA plan verification of 62 head and neck (HN) and 19 prostate cases was done using Octavius 4D setup with its associating VeriSoft gamma analysis software. The plans showed a maximum 3D gamma passing rate with 4 mm/3 % gamma acceptance criteria, i.e., 99.7 % for the HN cancer cases and 99.5 % for the prostate cancer cases. Local gamma analysis was also performed for both regions. Furthermore, 2D and volumetric gamma analyses were also carried out. Gamma analysis with respect to different axis was also carried out. It was known that the transversal axis showed    the highest gamma passing rate in both HN and prostate cases, i.e., 99.17 % and 98.3 %, respectively. The transverse axis came to be a better fit for the planned dose distribution.</dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US">NA</dc:contributor>
	<dc:date>2023-04-14</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1261</dc:identifier>
	<dc:identifier>10.55981/aij.2023.1261</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; VOL 49, NO 1 (2023): APRIL 2023; 53-59</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v1i1</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1261/1065</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1261/2271</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1261/2272</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1261/2273</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1261/2274</dc:relation>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>