A Multicenter Study of IMRT Dosimetry Audit Testing Using C-shape Phantom

Atom Indonesia · Vol. 51 No. 2 (2025) · hlm. 131–136

DOI
10.55981/aij.2025.1412
ISSN
2356-5322, 0126-1568
Terbit
2025-07-10
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

Intensity Modulation Radiation Therapy (IMRT) is a complex radiotherapy technique, so independent verification or dosimetry audits must be performed to ensure that accurate dosing is delivered to patients. This study conducted a multicenter audit using a dosimetry audit method developed from the IAEA dosimetry audit for IMRT/VMAT. The phantom in this study is made of acrylic material with two insert structures: planning target volume (PTV) and organ at risk (OAR). Phantom was scanned with a CT simulator at each hospital, and dose distribution was simulated with a PTV prescription dose of 4 Gy/2 fraction (D95 % = 95 %, D2 % 107 %, and Dmax 110 %) and a maximum OAR dose of 2.8 Gy. Dose evaluation in this study used TLD-rod for point dose and Gafchromic Film EBT3 for 2D dose distribution. Gamma evaluation was performed for film dose distribution with 3 %/3 mm and 3 %/2 mm criteria. The IMRT dosimetry audit using a C-shape phantom was tested in seven linacs (dynamic and static MLC) from six centers in Jakarta. The TLD results for PTV and OAR point dose show that all 14 IMRT plans meet deviation tolerance within ± 5 %. The film EBT3 evaluation identified that almost all plans pass the minimum 95 % gamma passing rate for 3 %/3 mm criteria and the minimum of 90 % for 3 %/2 mm. Three plans from three centers were also compared to the Gayatri (2022) study data from the same centers. Both results showed that all plans pass the action level ≥ 90 % for both 3 %/2 mm and 3 %/3 mm. Our audit dosimetry study approach employs a small and compact C-shaped phantom and dosimetry, facilitating easier distribution for remote audits. This study could serve as a starting point for remote audits leading to broader multicenter research in Indonesia.

Kata kunci

Jaringan kolaborasi para penulis

22 penulis · 65 tautan kepenulisan bersama · 17 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • University of Indonesia
  • Kementerian Kesehatan Republik Indonesia
  • Rumah Sakit Umum Pusat Nasional Dr. Cipto Mangunkusumo
  • Dharmais Cancer Hospital
  • National Nuclear Energy Agency of Indonesia
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
A AzziUniversity of Indonesia2
Djarwani Soeharso SoejokoUniversity of Indonesia2
W. PurwatiUniversity of Indonesia1
F. SuhaimiRumah Sakit Umum Pusat Nasional Dr. Cipto Mangunkusumo1
Wahyu Edy WibowoRumah Sakit Umum Pusat Nasional Dr. Cipto Mangunkusumo1
Dwi Seno Kuncoro SihonoUniversity of Indonesia1
R. D. SugandiUniversity of Indonesia1
V. VernandaUniversity of Indonesia1
Lukmanda Evan LubisUniversity of Indonesia1
Ika HariyatiAfiliasi tidak diketahui1
I A S Mu’minahUniversity of Indonesia1
T. MartUniversity of Indonesia1
K.Y.P. SandyNational Nuclear Energy Agency of Indonesia1
S. LiuraUniversity of Indonesia1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">A Multicenter Study of IMRT Dosimetry Audit Testing Using C-shape Phantom</dc:title>
	<dc:creator>Dwinesti, B.</dc:creator>
	<dc:creator>Ryangga, D.</dc:creator>
	<dc:creator>Dimitri, F.</dc:creator>
	<dc:creator>Handika, A. D.</dc:creator>
	<dc:creator>Fadli, M.</dc:creator>
	<dc:creator>Putranto, A. M. Y.</dc:creator>
	<dc:creator>Suharsono, S</dc:creator>
	<dc:creator>Pawiro, S. A.</dc:creator>
	<dc:subject xml:lang="en-US">IMRT, Dosimetry audit, Gamma analysis, TLD-rod, Film EBT3</dc:subject>
	<dc:description xml:lang="en-US">Intensity Modulation Radiation Therapy (IMRT) is a complex radiotherapy technique, so independent verification or dosimetry audits must be performed to ensure that accurate dosing is delivered to patients. This study conducted a multicenter audit using a dosimetry audit method developed from the IAEA dosimetry audit for IMRT/VMAT. The phantom in this study is made of acrylic material with two insert structures: planning target volume (PTV) and organ at risk (OAR). Phantom was scanned with a CT simulator at each hospital, and dose distribution was simulated with a PTV prescription dose of 4 Gy/2 fraction      (D95 % = 95 %, D2 %  107 %, and Dmax  110 %) and a maximum OAR dose of 2.8 Gy. Dose evaluation in this study used TLD-rod for point dose and Gafchromic Film EBT3 for 2D dose distribution. Gamma evaluation was performed for film dose distribution with 3 %/3 mm and 3 %/2 mm criteria. The IMRT dosimetry audit using a C-shape phantom was tested in seven linacs (dynamic and static MLC) from six centers in Jakarta. The TLD results for PTV and OAR point dose show that all 14 IMRT plans meet deviation tolerance within ± 5 %. The film EBT3 evaluation identified that almost all plans pass the minimum 95 % gamma passing rate for 3 %/3 mm criteria and the minimum of   90 % for 3 %/2 mm. Three plans from three centers were also compared to the Gayatri (2022) study data from the same centers. Both results showed that all plans pass the action level ≥ 90 % for both 3 %/2 mm and 3 %/3 mm.  Our audit dosimetry study approach employs a small and compact C-shaped phantom and dosimetry, facilitating easier distribution for remote audits. This study could serve as a starting point for remote audits leading to broader multicenter research in Indonesia.</dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US">Universitas Indonesia</dc:contributor>
	<dc:date>2025-07-10</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1412</dc:identifier>
	<dc:identifier>10.55981/aij.2025.1412</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; Vol 51, No 2 (2025): AUGUST 2025; 131-136</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v51i2</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1412/1174</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1412/2728</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1412/2729</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1412/2730</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2025 Atom Indonesia</dc:rights>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>