The Dose Distribution from Iridium-192 Source on Cervical Cancer Brachytherapy by Manchester System Using Monte Carlo Simulation

Atom Indonesia · Vol. 47 No. 3 (2021) · hlm. 205–211

DOI
10.17146/aij.2021.1102
ISSN
2356-5322, 0126-1568
Terbit
2021-12-03
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

One treatment for cervical cancer is to use radioactive sources that directly target the cancer cell called brachytherapy. This study is aimed to determine dose distribution at phantom pelvis using the DOSXYZnrc Monte Carlo code. The phantom was derived from a CT scan image of the DICOM-type pelvis with a size of 50 × 50 × 28.8 cm obtained from Santosa Kopo Hospital. The source used was Ir-192, which makes an asymmetrical beam with a size of 0.45 × 0.09 × 0.09 cm. Monte Carlo simulation was performed to determine the dose distribution of the Ir-192 source on cervical cancer CT images based on the Manchester system. The Monte Carlo simulation was divided into two models with distance variations on the applicator. Model A used TPS data with a distance between sources of 0.9 cm, while model B had a distance between sources of 0.5 cm. The distribution of dose resulting from the Monte Carlo simulation was analyzed and compared with TPS data. The results showed that at the range of 50 %, dose distribution in model A reaches the end of 3.9 cm. When compared to the range of 50 % dose distribution at the TPS results that reaches the point of 4 cm, it produces a deviation value of 2.5 %, which is still within the tolerance range. Model A and Model B provide different dose distribution. In model B, it reaches 3.86 cm, resulting in a deviation of 1.02 %, which is still within the tolerance range. The resulting γ-index value for the 50 % dose distribution was 2.26, while the whole area's GPR value was 94.13 %. This indicates a difference in dose distribution between the two models. Therefore, the smaller the distance between the sources, the shorter the dose distribution range with relatively more uniform dose distribution.

Kata kunci

Jaringan kolaborasi para penulis

25 penulis · 73 tautan kepenulisan bersama · 51 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Bandung Institute of Technology
  • Universitas Matana
  • Sunan Kalijaga State Islamic University Yogyakarta
  • Syarif Hidayatullah State Islamic University Jakarta
  • Universitas Prima Nusantara Bukittinggi
  • Diponegoro University
  • California State University, Channel Islands
  • Organisasi lain
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Choirul AnamDiponegoro University2
Josua Timotius ManikUniversitas Matana2
Anisza OkseliaDr. Hasan Sadikin General Hospital2
Angelica MargarethaUniversitas Matana2
Rena WiditaBandung Institute of Technology1
Idam ArifBandung Institute of Technology1
Geoff DoughertyCalifornia State University, Channel Islands1
Mohamad Fahdillah RhaniTan Tock Seng Hospital1
S. YaniIPB University1
Nita HandayaniSunan Kalijaga State Islamic University Yogyakarta1
Warsito Purwo TarunoAfiliasi tidak diketahui1
Kharisma Fajar HidayantiSunan Kalijaga State Islamic University Yogyakarta1
Siti Nurul KhotimahBandung Institute of Technology1
Marlin Ramadhan BaidillahAfiliasi tidak diketahui1
Mitra DjamalBandung Institute of Technology1
Ramli RamliUniversitas Negeri Padang1
Ahmad AminudinIndonesia University of Education1
Khairurrijal KhairurrijalBandung Institute of Technology1
Ambran HartonoSyarif Hidayatullah State Islamic University Jakarta1
Edi SanjayaSyarif Hidayatullah State Islamic University Jakarta1
Cuk ImawanUniversity of Indonesia1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">The Dose Distribution from Iridium-192 Source on  Cervical Cancer Brachytherapy by Manchester  System Using Monte Carlo Simulation</dc:title>
	<dc:creator>Kurniati, F.</dc:creator>
	<dc:creator>Krisna, F. P.</dc:creator>
	<dc:creator>Junios, J.</dc:creator>
	<dc:creator>Haryanto, F.</dc:creator>
	<dc:subject xml:lang="en-US">Brachytherapy; CT image; Ir-192 source; Manchester system; Monte carlo simulation</dc:subject>
	<dc:description xml:lang="en-US">One treatment for cervical cancer is to use radioactive sources that directly target the cancer cell called brachytherapy. This study is aimed to determine dose distribution at phantom pelvis using the DOSXYZnrc Monte Carlo code. The phantom was derived from a CT scan image of the DICOM-type pelvis with a size of 50 × 50 × 28.8 cm obtained from Santosa Kopo Hospital. The source used was Ir-192, which makes an asymmetrical beam with a size of 0.45 × 0.09 × 0.09 cm. Monte Carlo simulation was performed to determine the dose distribution of the Ir-192 source on cervical cancer CT images based on the Manchester system. The Monte Carlo simulation was divided into two models with distance variations on the applicator. Model A used TPS data with a distance between sources of 0.9 cm, while model B had a distance between sources of 0.5 cm. The distribution of dose resulting from the Monte Carlo simulation was analyzed and compared with TPS data. The results showed that at the range of 50 %, dose distribution in model A reaches the end of 3.9 cm. When compared to the range of 50 % dose distribution at the TPS results that reaches the point of 4 cm, it produces a deviation value of 2.5 %, which is still within the tolerance range. Model A and Model B provide different dose distribution. In model B, it reaches 3.86 cm, resulting in a deviation of 1.02 %, which is still within the tolerance range. The resulting γ-index value for the 50 % dose distribution was 2.26, while the whole area's GPR value was 94.13 %. This indicates a difference in dose distribution between the two models. Therefore, the smaller the distance between the sources, the shorter the dose distribution range with relatively more uniform dose distribution. </dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US">This study was fully supported by Program Penelitian Kolaborasi Indonesia (PPKI) (grant number 020/WCU-ITB/LL/I/2020)</dc:contributor>
	<dc:date>2021-12-03</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1102</dc:identifier>
	<dc:identifier>10.17146/aij.2021.1102</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; Vol 47, No 3 (2021): December 2021; 205-211</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v47i3</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1102/996</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1102/1908</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1102/2077</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1102/2078</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1102/2079</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1102/2080</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2021 Atom Indonesia</dc:rights>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>