Estimating TPR₂₀,₁₀ Under Non-Reference Conditions Using a Geometric Sequence Approach

Atom Indonesia · Vol. 52 No. 1 (2026) · hlm. 73–79

DOI
10.55981/aij.2026.1564
ISSN
2356-5322, 0126-1568
Terbit
2026-03-31
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

An alternative approach to estimate the Tissue Phantom Ratio (TPR) at depths of 20 cm and 10 cm (TPR₂₀,₁₀) under non-reference conditions is required to address situations where a 10 × 10 cm² field size is not achievable on a specific Linear Accelerator (LINAC) during a beam quality test. This study aims to estimate the TPR20,10 under non-reference conditions using a geometric sequence approach, and to compare it with the TPR₂₀,₁₀ under non-reference conditions estimated using the Sauer method, the Palmas method, a linear fit approach, as well as with the TPR₂₀,₁₀ under reference conditions calculated using the TRS-398 protocol. The first step in this study was measuring the percentage depth dose (PDD), D20cm, and D10cm with field size variations from 4 × 4 cm² to 10 × 10 cm² for both 6 MV and 10 MV X-ray beams. The PDD were used to estimate the TPR₂₀,₁₀ using a geometric sequence approach, the Sauer method, the Palmans method, and a linear fit approach, and to calculate the TPR₂₀,₁₀ using the TRS-398 protocol. The D20cm and D10cm were also used to calculate the TPR₂₀,₁₀ using the TRS-398 protocol. The TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence approach were 0.683 ± 0.004 and 0.742 ± 0.005, respectively. The level of precision that could be reached by the geometric sequence approach is potentially equivalent to the TRS-398 protocol, the Sauer method, the Palmans method, and the linear fit approach. The TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence method did not show a significant difference compared with the TPR₂₀,₁₀ calculated using the TRS-398 protocol. However, the TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence approach showed a significant difference compared with those TPR₂₀,₁₀ estimated using the Sauer method and the Palmans method.

Kata kunci

Jaringan kolaborasi para penulis

18 penulis · 48 tautan kepenulisan bersama

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • University of Brawijaya
  • Mulawarman University
  • RSUD Prof. Dr. Saiful Anwar Malang
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Sri HerwiningsihUniversity of Brawijaya3
Johan Andoyo Effendi NoorUniversity of Brawijaya2
D. Y. B. MuntheUniversity of Brawijaya2
Alfian K AntoAfiliasi tidak diketahui1
Rafiq Sulistyo NugrohoRSUD Prof. Dr. Saiful Anwar Malang1
A. HidayatUniversity of Brawijaya1
Firdy YuanaUniversity of Brawijaya1
Riska Nurul LatifahUniversity of Brawijaya1
D. P. RahmahtullahUniversity of Brawijaya1
I. AlvianiUniversity of Brawijaya1
Erlinda Ratnasari PutriMulawarman University1
D. R. (Dian) PutriMulawarman University1
Robert Janssen StevenlyAfiliasi tidak diketahui1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Estimating TPR₂₀,₁₀ Under Non-Reference Conditions Using a Geometric Sequence Approach</dc:title>
	<dc:creator>Rizaldi, F. A.</dc:creator>
	<dc:creator>Herwiningsih, S.</dc:creator>
	<dc:creator>Widodo, C. S.</dc:creator>
	<dc:creator>Hentihu, F. K.</dc:creator>
	<dc:creator>Anto, A. K.</dc:creator>
	<dc:subject xml:lang="en-US">TPR20,10 under reference conditions; TPR20,10 under non-reference conditions; Beam quality test of linear accelerator; Geometric sequence approach; X-ray megavoltage photon beam;</dc:subject>
	<dc:description xml:lang="en-US">An alternative approach to estimate the Tissue Phantom Ratio (TPR) at depths of 20 cm and 10 cm (TPR₂₀,₁₀) under non-reference conditions is required to   address situations where a 10 × 10 cm² field size is not achievable on a specific Linear Accelerator (LINAC) during a beam quality test. This study aims to estimate the TPR20,10 under non-reference conditions using a geometric sequence approach, and to compare it with the TPR₂₀,₁₀ under non-reference conditions estimated using the Sauer method, the Palmas method, a linear fit approach, as well as with the TPR₂₀,₁₀ under reference conditions calculated using the TRS-398 protocol. The first step in this study was measuring the percentage depth dose (PDD), D20cm, and D10cm with field size variations from 4 × 4 cm² to 10 × 10 cm² for both 6 MV and 10 MV X-ray beams. The PDD were used to estimate the TPR₂₀,₁₀ using a geometric sequence approach, the Sauer method, the Palmans method, and a linear fit approach, and to calculate the TPR₂₀,₁₀ using the TRS-398 protocol. The D20cm and D10cm were also used to calculate the TPR₂₀,₁₀ using the TRS-398 protocol. The TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence approach were 0.683 ± 0.004 and 0.742 ± 0.005, respectively. The level of precision that could be reached by the geometric sequence approach is potentially equivalent to the TRS-398 protocol, the Sauer method, the Palmans method, and the linear fit approach. The TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence method did not show a significant difference compared with the TPR₂₀,₁₀ calculated using the TRS-398 protocol. However, the TPR₂₀,₁₀ for 6 MV and 10 MV X-ray beams estimated using the geometric sequence approach showed a significant difference compared with those TPR₂₀,₁₀ estimated using the Sauer method and the Palmans method.</dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US">Brawijaya University, Physics Department</dc:contributor>
	<dc:contributor xml:lang="en-US">Lavalette Hospital, Radiotherapy Department</dc:contributor>
	<dc:date>2026-03-31</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1564</dc:identifier>
	<dc:identifier>10.55981/aij.2026.1564</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; Vol 52, No 1 (2026): APRIL 2026; 73-79</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v52i1</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1564/1210</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2557</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2558</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2603</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2678</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2679</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2938</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2939</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2940</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1564/2941</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2026 Atom Indonesia</dc:rights>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>