Dose Response of Hp(0.07) on TLD-700 and TLD-900 Based Ring Dosimeters to the Sr-90 Exposure

Atom Indonesia · Vol. 50 No. 3 (2024) · hlm. 267–271

DOI
10.55981/aij.2024.1425
ISSN
2356-5322, 0126-1568
Terbit
2024-11-22
Penerbit
National Research and Innovation Agency
Bahasa
eng
Lisensi
creativecommons.org/licenses/by-nc-sa/4.0

Abstrak

The preparation of radiopharmaceuticals may necessitate close contact with the radioactive source. In this case, high-energy beta-emitting nuclides can cause a high exposure. There are two methods of measuring the dose in the extremities, including using a bracelet and a ring dosimeter. In this study, the response of Hp(0.07) dose in TLD-700 and TLD-900-based ring dosimeters to Sr-90 beta radiation will be compared for parameters of distance, time, and angle of irradiation. The dosimeters were irradiated using Sr-90 at a distance of 5, 7, and 10 cm from the source surface and all of them were subsequently read with TLD Reader. The same type of dosimeters were exposed to Sr-90 for 1 day, 2 days, and 3 days at a distance of 7 cm and read. At a distance of 7 cm, the other dosimeters were irradiated at angles of 30°, 60°, -30°, and -60°, and then read. The study shows that the relationship between the distance of irradiation to Hp(0.07) dose is polynomial with a correlation factor of 1, both for TLD-700 and TLD-900-based ring dosimeters. The dose response at the variation of irradiation time showed linear behavior with correlation coefficients of 0.9942 and 0.9999 for TLD-700 and TLD-900, respectively. The decrease in Hp(0.07) response in TLD-700 reached 14-31 % for the irradiation angle of ±30o and 23-27 % in TLD-900 for the same angle. Meanwhile, the decrease in Hp(0.07) response at the angle of irradiation ±60o reached 48 % for TLD-700 and 66-67 % for TLD-900.

Kata kunci

Jaringan kolaborasi para penulis

13 penulis · 41 tautan kepenulisan bersama · 7 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • National Research and Innovation Agency
  • National Nuclear Energy Agency of Indonesia
  • Rumah Sakit Umum Pusat Sanglah Denpasar
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Rofiq SyaifudinNational Research and Innovation Agency3
Okky Agassy FirmansyahNational Nuclear Energy Agency of Indonesia1
A. F. FirmansyahNational Nuclear Energy Agency of Indonesia1
Sunaryati Sri InangNational Nuclear Energy Agency of Indonesia1
A. R. SetiadiAfiliasi tidak diketahui1
O. A. AkbarAfiliasi tidak diketahui1
V. ArifAfiliasi tidak diketahui1
C. AmeliaRumah Sakit Umum Pusat Sanglah Denpasar1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://atomindonesia.brin.go.id/index.php/aij/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 8 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Dose Response of Hp(0.07) on TLD-700 and TLD-900 Based Ring Dosimeters to the Sr-90 Exposure</dc:title>
	<dc:creator>Jumpeno, E. B.</dc:creator>
	<dc:creator>Anggraeni, R.</dc:creator>
	<dc:creator>Putri, M. Muharani</dc:creator>
	<dc:creator>Barokah, S. N.</dc:creator>
	<dc:creator>Adisti, F. N.</dc:creator>
	<dc:subject xml:lang="en-US">Hp(0.07); TLD-700;  TLD-900; Ring dosimeter; Sr-90</dc:subject>
	<dc:description xml:lang="en-US">The preparation of radiopharmaceuticals may necessitate close contact with the radioactive source. In this case, high-energy beta-emitting nuclides can cause a high exposure. There are two methods of measuring the dose in the extremities, including using a bracelet and a ring dosimeter. In this study, the response of Hp(0.07) dose in TLD-700 and TLD-900-based ring dosimeters to Sr-90 beta radiation will be compared for parameters of distance, time, and angle of irradiation. The dosimeters were irradiated using Sr-90 at a distance of 5, 7, and  10 cm from the source surface and all of them were subsequently read with TLD Reader. The same type of dosimeters were exposed to Sr-90 for 1 day, 2 days, and 3 days at a distance of 7 cm and read. At a distance of 7 cm, the other dosimeters were irradiated at angles of 30°, 60°, -30°, and -60°, and then read. The study shows that the relationship between the distance of irradiation to Hp(0.07) dose is polynomial with a correlation factor of 1, both for TLD-700 and TLD-900-based ring dosimeters. The dose response at the variation of irradiation time showed linear behavior with correlation coefficients of 0.9942 and 0.9999 for TLD-700 and TLD-900, respectively. The decrease in Hp(0.07) response in TLD-700 reached 14-31 % for the irradiation angle of ±30o and 23-27 % in TLD-900 for the same angle. Meanwhile, the decrease in Hp(0.07) response at the angle of irradiation ±60o reached 48 % for TLD-700 and 66-67 % for TLD-900.</dc:description>
	<dc:publisher xml:lang="en-US">National Research and Innovation Agency</dc:publisher>
	<dc:contributor xml:lang="en-US">National Research and Inovation Agency and the General Manager Dosimetry Laboratory of NuklindoLab</dc:contributor>
	<dc:date>2024-11-22</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en-US"></dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://atomindonesia.brin.go.id/index.php/aij/article/view/1425</dc:identifier>
	<dc:identifier>10.55981/aij.2024.1425</dc:identifier>
	<dc:source xml:lang="en-US">Atom Indonesia; Vol 50, No 3 (2024): DECEMBER 2024; 267-271</dc:source>
	<dc:source>2356-5322</dc:source>
	<dc:source>0126-1568</dc:source>
	<dc:source>10.55981/aij.v50i3</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/view/1425/1141</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2562</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2563</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2564</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2565</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2568</dc:relation>
	<dc:relation>https://atomindonesia.brin.go.id/index.php/aij/article/downloadSuppFile/1425/2579</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2024 Atom Indonesia</dc:rights>
	<dc:rights xml:lang="en-US">http://creativecommons.org/licenses/by-nc-sa/4.0</dc:rights>
</oai_dc:dc>