Are mechanical and electromechanical methods accurately interchangeable for measuring plasma prothrombin time and activated partial thromboplastin time? A comparative analysis study and potential implication to cardiovascular disease

Heart Science Journal · Vol. 6 No. 1 (2025) · hlm. 72–76

DOI
10.21776/ub.hsj.2025.006.01.12
ISSN
2721-9984, 2721-9976
Terbit
2025-01-31
Penerbit
Universitas Brawijaya
Bahasa
en
Lisensi
creativecommons.org/licenses/by/4.0

Abstrak

INTRODUCTION: The DT100 offers both optical and mechanical modes, with its mechanical mode showing better homogenization than the STAGO, but comparative study is limited. OBJECTIVES: The study aimed to evaluate the diagnostic accuracy of plasma Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) measurements using the DT100 and STAGO instruments. METHODS: Designated as a cross-sectional study, this study was conducted at RSUD Dr. Soetomo from October 2022 to January 2023. Venous blood samples with plasma citrate anticoagulant 0.109 M 3.2% were consecutively collected from hospitalized patients, and all samples underwent testing using both the DT100 in mechanical mode (DT100, TCoag Ireland Limited, Ireland) and the STAGO employing an electromechanical method (Compact Max3, STAGO, France). Statistical analysis included comparison using Paired t-test, Pearson correlation, and Bland-Altman analysis to assess agreement between the results obtained from the two instruments. RESULTS: The study included 51 patients. PT levels were significantly lower with the DT100 compared to STAGO (MD: -2.0; 95%CI: (-2.30) – (-1.3); p<0.0001), and showed a strong positive correlation between methods (r:0.9535; p<0.0001). However, Bland-Altman analysis for PT showed a bias of 1.84, with limits of agreement (3.30-0.37), indicating systematic differences and variability. APTT levels were significantly higher with DT100 compared to STAGO (MD:3.60; 95%CI: 2.13–5.07; p<0.0001), with a moderate positive correlation (r:0.6690; p<0.0001). For APTT, bias of Bland-Altman analysis was -3.60, with limits ((-9.84) – (2.64)), suggesting significant discrepancies and variability between methods. CONCLUSION: The study found significant variability in PT and APTT measurements between the DT100 and STAGO methods.

Kata kunci

Jaringan kolaborasi para penulis

5 penulis · 6 tautan kepenulisan bersama

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Universitas Airlangga
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Nur Aisiyah WidjajaUniversitas Airlangga1
Gwyneth Trixie StanpoUniversitas Airlangga1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://heartscience.ub.ac.id/index.php/heartscience/oai, pertama di-harvest 8 Nov 2026, terakhir terlihat 8 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Are mechanical and electromechanical methods accurately interchangeable for measuring plasma prothrombin time and activated partial thromboplastin time? A comparative analysis study and potential implication to cardiovascular disease</dc:title>
	<dc:creator xml:lang="en">Pranata, Fransiska Jaya</dc:creator>
	<dc:creator xml:lang="en">Notopuro, Paulus Budiono</dc:creator>
	<dc:creator xml:lang="en">Hajat, Arifoel</dc:creator>
	<dc:subject xml:lang="en">Prothrombin time</dc:subject>
	<dc:subject xml:lang="en">activated partial thromboplastin time</dc:subject>
	<dc:subject xml:lang="en">mechanical method</dc:subject>
	<dc:subject xml:lang="en">DT100 TCoag</dc:subject>
	<dc:subject xml:lang="en">STAGO Compact Max 3</dc:subject>
	<dc:description xml:lang="en">INTRODUCTION: The DT100 offers both optical and mechanical modes, with its mechanical mode showing better homogenization than the STAGO, but comparative study is limited.
OBJECTIVES: The study aimed to evaluate the diagnostic accuracy of plasma Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) measurements using the DT100 and STAGO instruments.
METHODS: Designated as a cross-sectional study, this study was conducted at RSUD Dr. Soetomo from October 2022 to January 2023. Venous blood samples with plasma citrate anticoagulant 0.109 M 3.2% were consecutively collected from hospitalized patients, and all samples underwent testing using both the DT100 in mechanical mode (DT100, TCoag Ireland Limited, Ireland) and the STAGO employing an electromechanical method (Compact Max3, STAGO, France). Statistical analysis included comparison using Paired t-test, Pearson correlation, and Bland-Altman analysis to assess agreement between the results obtained from the two instruments.
RESULTS: The study included 51 patients. PT levels were significantly lower with the DT100 compared to STAGO (MD: -2.0; 95%CI: (-2.30) – (-1.3); p&amp;lt;0.0001), and showed a strong positive correlation between methods (r:0.9535; p&amp;lt;0.0001). However, Bland-Altman analysis for PT showed a bias of 1.84, with limits of agreement (3.30-0.37), indicating systematic differences and variability. APTT levels were significantly higher with DT100 compared to STAGO (MD:3.60; 95%CI: 2.13–5.07; p&amp;lt;0.0001), with a moderate positive correlation (r:0.6690; p&amp;lt;0.0001). For APTT, bias of Bland-Altman analysis was -3.60, with limits ((-9.84) – (2.64)), suggesting significant discrepancies and variability between methods.
CONCLUSION: The study found significant variability in PT and APTT measurements between the DT100 and STAGO methods.</dc:description>
	<dc:publisher xml:lang="en">Universitas Brawijaya</dc:publisher>
	<dc:date>2025-01-31</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://heartscience.ub.ac.id/index.php/heartscience/article/view/20186</dc:identifier>
	<dc:identifier>10.21776/ub.hsj.2025.006.01.12</dc:identifier>
	<dc:source xml:lang="en">Heart Science Journal; Vol. 6 No. 1 (2025): Challenges in Managing Acute Heart Failure; 72-76</dc:source>
	<dc:source>2721-9984</dc:source>
	<dc:source>2721-9976</dc:source>
	<dc:source>10.21776/ub.hsj.2025.006.01</dc:source>
	<dc:language>en</dc:language>
	<dc:relation>https://heartscience.ub.ac.id/index.php/heartscience/article/view/20186/472</dc:relation>
	<dc:relation>10.21776/</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2025 Fransiska Jaya Pranata, Paulus Budiono Notopuro, Arifoel Hajat</dc:rights>
	<dc:rights xml:lang="en">http://creativecommons.org/licenses/by/4.0</dc:rights>
</oai_dc:dc>