Impact of Different Grids on the Quality of Abdominal Radiography Images

JURNAL ILMIAH PENDIDIKAN FISIKA AL BIRUNI · Vol. 13 No. 1 (2024) · hlm. 95–105

DOI
10.24042/jipfalbiruni.v13i1.22202
ISSN
2503-023X, 2303-1832
Terbit
2024-06-29
Penerbit
Universitas Islam Negeri Raden Intan Lampung, Indonesia
Bahasa
eng

Abstrak

Grids are additional accessory in radiography that absorb scatter radiation hoping to reduce blurriness on the image. Various type of grids is available in the market so the effect of those on image quality needs to be evaluated. The purpose of this study is to compare the quality of abdominal radiography images using three variations of grids (moving and stationary grid) and without using a grid in order to see the best type or grids to produce the greatest image quality. Abdominal radiography exposure factors were used to expose two phantoms. Pro-Fluo 150 phantom was used as the object to measure image quality such as Signal to Noise Ratio (SNR), Contrast to Noise Ratio (CNR), and spatial resolution. A survey of five radiologists was also performed to evaluate the contrast and the visibility of objects’ boundaries in Anthropomorphic phantom images. The result of this study showed that the usage of various grids gave different values of SNR, CNR, and spatial resolution. The exposure factor of 70-75 kV in each mAs have the highest SNR value (52,64-78,31). Images obtained without a grid and exposure factor of 80-85 kV in each mAs have the highest SNR value (160,65-800,48). Images obtained from using moving grid also have the highest CNR and spatial resolution, with score of 3.72 – 7.62 and 3.7 lp/mm. The results of radiologists’ survey showed that the average score for moving grid’s images was 29.8 of 30 with a percentage of 99.33%. Therefore, it that can concluded that images obtained with moving grid have the best quality that can provides contrast and shows the boundaries between organs clearly in abdominal radiography. Based on this research, it’s recommended to use moving grid for abdominal radiography examinations.

Kata kunci

Jaringan kolaborasi para penulis

9 penulis · 19 tautan kepenulisan bersama · 8 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Nasional University
  • State University of Jakarta
  • National Chiayi University
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Deni Tiko BahagiaNasional University3
Dadan SumardaniNational Chiayi University1
Fauzi BakriState University of Jakarta1
Agustiani PutriState University of Jakarta1
Rahma Rosaliana SaraswatiState University of Jakarta1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) http://ejournal.radenintan.ac.id/index.php/al-biruni/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Impact of Different Grids on the Quality of Abdominal Radiography Images</dc:title>
	<dc:creator xml:lang="en">Shekti, Shitoch Armandho Bimo</dc:creator>
	<dc:creator xml:lang="en">Sari, Ni Larasati Kartika</dc:creator>
	<dc:creator xml:lang="en">Hartoyo, Puji</dc:creator>
	<dc:creator xml:lang="en">Muliyati, Dewi</dc:creator>
	<dc:subject xml:lang="en">Noise Ratio</dc:subject>
	<dc:subject xml:lang="en">Signal to Noise Ratio</dc:subject>
	<dc:subject xml:lang="en">Spatial Resolution</dc:subject>
	<dc:description xml:lang="en">Grids are additional accessory in radiography that absorb scatter radiation hoping to reduce blurriness on the image.  Various type of grids is available in the market so the effect of those on image quality needs to be evaluated. The purpose of this study is to compare the quality of abdominal radiography images using three variations of grids (moving and stationary grid) and without using a grid in order to see the best type or grids to produce the greatest image quality. Abdominal radiography exposure factors were used to expose two phantoms. Pro-Fluo 150 phantom was used as the object to measure image quality such as Signal to Noise Ratio (SNR), Contrast to Noise Ratio (CNR), and spatial resolution. A survey of five radiologists was also performed to evaluate the contrast and the visibility of objects’ boundaries in Anthropomorphic phantom images. The result of this study showed that the usage of various grids gave different values of SNR, CNR, and spatial resolution. The exposure factor of 70-75 kV in each mAs have the highest SNR value (52,64-78,31). Images obtained without a grid and exposure factor of 80-85 kV in each mAs have the highest SNR value (160,65-800,48). Images obtained from using moving grid also have the highest CNR and spatial resolution, with score of 3.72 – 7.62 and 3.7 lp/mm. The results of radiologists’ survey showed that the average score for moving grid’s images was 29.8 of 30 with a percentage of 99.33%. Therefore, it that can concluded that images obtained with moving grid have the best quality that can provides contrast and shows the boundaries between organs clearly in abdominal radiography. Based on this research, it’s recommended to use moving grid for abdominal radiography examinations.</dc:description>
	<dc:publisher xml:lang="en">Universitas Islam Negeri Raden Intan Lampung, Indonesia</dc:publisher>
	<dc:date>2024-06-29</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en">Peer-reviewed Article</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/22202</dc:identifier>
	<dc:identifier>10.24042/jipfalbiruni.v13i1.22202</dc:identifier>
	<dc:source xml:lang="en">Jurnal Ilmiah Pendidikan Fisika Al-Biruni; Vol. 13 No. 1 (2024): Jurnal Ilmiah Pendidikan Fisika Al-Biruni; 95-105</dc:source>
	<dc:source xml:lang="id">Jurnal Ilmiah Pendidikan Fisika Al-Biruni; Vol 13 No 1 (2024): Jurnal Ilmiah Pendidikan Fisika Al-Biruni; 95-105</dc:source>
	<dc:source>2503-023X</dc:source>
	<dc:source>2303-1832</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/22202/pdf</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2024 Physics Education Department, UIN Raden Intan Lampung, Indonesia</dc:rights>
</oai_dc:dc>