Finite Element Analysis of SS316L-Based Five-Hole Plate Implant For Fibula Reconstruction

Journal of Applied Engineering and Technological Science (JAETS) · Vol. 4 No. 1 (2022) · hlm. 16–23

DOI
10.37385/jaets.v4i1.533
ISSN
2715-6079, 2715-6087
Terbit
2022-08-15
Penerbit
Yayasan Riset dan Pengembangan Intelektual (YRPI)
Bahasa
en

Abstrak

This study analyzed the design performance of SS316L-based plate implant for fibula restoration using a Finite Element Analysis approach. The simulated model design has dimensions of 35 x 5 x 1.5 mm and five holes with 2-3 configuration. The results of the bending test simulation showed that the values for both displacement and Von Mises stress that occurred (0.008 mm and 116 MPa of each) were still considerably below the yield stress of the SS316L material. The same results were also shown in the tensile test simulation, although the clamping setting on the plate was changed on the other side. From this finite element analysis approach, the SS316L-based five-hole plate implant design has a fairly good strength performance as a fibular bone-implant restoration product.

Kata kunci

Jaringan kolaborasi para penulis

20 penulis · 62 tautan kepenulisan bersama · 32 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Politeknik Negeri Malang
  • Universitas Negeri Padang
  • State University of Malang
  • Politeknik Negeri Samarinda
  • Cranfield University
  • Politeknik Negeri Semarang
  • Politeknik Negeri Banyuwangi
  • Organisasi lain
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Ali Sai’inPoliteknik Negeri Semarang7
Andre KurniawanUniversitas Negeri Padang4
Fina Andika Frida AstutiPoliteknik Negeri Malang4
Ahmad Hanif FirdausPoliteknik Negeri Malang4
Lutfi HakimPoliteknik Negeri Banyuwangi3
Moh. HartonoPoliteknik Negeri Malang3
Moh. Nasir HariyantoPoliteknik Negeri Malang3
Syamsul HadiUniversity of Sarjanawiyata Tamansiswa3
Poppy PuspitasariState University of Malang1
Putut MurdantoState University of Malang1
Mohammad Sugi HartonoPoliteknik Negeri Malang1
Remon LapisaUniversitas Negeri Padang1
Jasman JasmanUniversitas Negeri Padang1
Zainal AbadiUniversitas Negeri Padang1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://journal.yrpipku.com/index.php/jaets/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 9 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Finite Element Analysis of SS316L-Based Five-Hole Plate Implant For Fibula Reconstruction</dc:title>
	<dc:creator xml:lang="en">Qosim, Nanang</dc:creator>
	<dc:creator xml:lang="en">Emzain, Zakki Fuadi</dc:creator>
	<dc:creator xml:lang="en">Mufarrih, AM.</dc:creator>
	<dc:creator xml:lang="en">Monasari, Ratna</dc:creator>
	<dc:creator xml:lang="en">Kusumattaqiin, Fataa</dc:creator>
	<dc:creator xml:lang="en">Santoso, Rangga E.</dc:creator>
	<dc:subject xml:lang="en">Plate</dc:subject>
	<dc:subject xml:lang="en">Implant</dc:subject>
	<dc:subject xml:lang="en">Finite Element Method</dc:subject>
	<dc:subject xml:lang="en">Von Mises</dc:subject>
	<dc:subject xml:lang="en">Fibula</dc:subject>
	<dc:subject xml:lang="en">SS316L</dc:subject>
	<dc:description xml:lang="en">This study analyzed the design performance of SS316L-based plate implant for fibula restoration using a Finite Element Analysis approach. The simulated model design has dimensions of 35 x 5 x 1.5 mm and five holes with 2-3 configuration. The results of the bending test simulation showed that the values for both displacement and Von Mises stress that occurred (0.008 mm and 116 MPa of each) were still considerably below the yield stress of the SS316L material. The same results were also shown in the tensile test simulation, although the clamping setting on the plate was changed on the other side. From this finite element analysis approach, the SS316L-based five-hole plate implant design has a fairly good strength performance as a fibular bone-implant restoration product.</dc:description>
	<dc:publisher xml:lang="en">Yayasan Riset dan Pengembangan Intelektual (YRPI)</dc:publisher>
	<dc:date>2022-08-15</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.yrpipku.com/index.php/jaets/article/view/533</dc:identifier>
	<dc:identifier>10.37385/jaets.v4i1.533</dc:identifier>
	<dc:source xml:lang="en">Journal of Applied Engineering and Technological Science (JAETS); Vol. 4 No. 1 (2022): Journal of Applied Engineering and Technological Science (JAETS); 16-23</dc:source>
	<dc:source>2715-6079</dc:source>
	<dc:source>2715-6087</dc:source>
	<dc:source>10.37385/jaets.v4i1</dc:source>
	<dc:language>en</dc:language>
	<dc:relation>https://journal.yrpipku.com/index.php/jaets/article/view/533/563</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2022 Journal of Applied Engineering and Technological Science (JAETS)</dc:rights>
</oai_dc:dc>