Noise Reduction of Motion and EMG Artifacts in Holter ECG Using IIR Filters for Robust Arrhythmia Detection

Journal of Applied Engineering and Technological Science (JAETS) · Vol. 7 No. 2 (2026) · hlm. 1177–1201

DOI
10.37385/jaets.v7i2.9426
ISSN
2715-6079, 2715-6087
Terbit
2026-06-15
Penerbit
Yayasan Riset dan Pengembangan Intelektual (YRPI)
Bahasa
en

Abstrak

Ambulatory Holter electrocardiography (ECG) enables continuous monitoring for detecting transient arrhythmias; however, its diagnostic reliability is significantly degraded by motion artifacts and electromyographic (EMG) interference. Under severe motion artifact conditions, prior studies report that ambulatory ECG SNR can fall below −10 dB, although SNR levels vary substantially depending on activity type and electrode placement, reducing usable data segments and impairing arrhythmia detection. While advanced denoising methods such as wavelet transforms and deep learning achieve high accuracy, their computational complexity limits real-time deployment in resource-constrained embedded systems. This reveals a critical gap in lightweight methods that jointly optimize noise suppression, morphological preservation, and downstream diagnostic performance. This study proposes a computationally efficient IIR Butterworth bandpass filtering framework for real-time IoT-based Holter ECG systems. The system combines three-lead ECG acquisition, embedded processing on an ESP32, and real-time visualization. Performance is assessed using SNR, mean squared error (MSE), Pearson correlation, and confusion matrix-based detection metrics on ten male participants under controlled motion and muscle artifact conditions. Results demonstrate statistically significant SNR improvements for motion artifacts (ΔSNR = 9.47 ± 1.96 dB, t(9) = 15.28, p < 0.001) and EMG artifacts (ΔSNR = 16.73 ± 0.91 dB, t(9) = 58.11, p < 0.0001). Post-filtering morphological fidelity was high, with mean Pearson correlation of 0.963 for motion artifacts and 0.945 for muscle artifacts. These signal quality improvements translated into 95.3% post-filtering arrhythmia detection accuracy (sensitivity: ≈96.0%, specificity: ≥97.0%, F1-score: ≥95.0%), significantly exceeding the 70% minimum performance threshold adopted in this study as a conservative screening criterion (t(9) = 29.7, p < 0.001). Despite dataset limitations (n = 10), the proposed framework provides an effective trade-off between computational efficiency and diagnostic reliability, supporting scalable and real-time ambulatory ECG monitoring for early arrhythmia screening.

Kata kunci

Jaringan kolaborasi para penulis

23 penulis · 58 tautan kepenulisan bersama · 19 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Poltekkes Kemenkes Surabaya
  • Universiti Brunei Darussalam
  • Bani-Waleed University
  • Universitas Airlangga
  • Rumah Sakit Dokter Soetomo
  • Poltekkes Kemenkes Malang
  • Azerbaijan State Oil and Industry University
  • Organisasi lain
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Wahyu CaesarendraUniversiti Brunei Darussalam2
Sari LuthfiyahPoltekkes Kemenkes Surabaya2
Abdussalam Ali AhmedBani-Waleed University2
Endro YuliantoPoltekkes Kemenkes Surabaya1
Umaimah Mitsalia Ummi SalwaUniversitas Airlangga1
Tri Bowo IndartoPoltekkes Kemenkes Surabaya1
Triana RahmawatiAfiliasi tidak diketahui1
I Putu Alit PawanaRumah Sakit Dokter Soetomo1
Bedjo UtomoPoltekkes Kemenkes Surabaya1
Sugeng Iwan SetyobudiPoltekkes Kemenkes Malang1
Herianto HeriantoUniversitas Gadjah Mada1
Vugar Hajimahmud AbdullayevAzerbaijan State Oil and Industry University1
Vugar AbdullayevAfiliasi tidak diketahui1
Abd KholiqAfiliasi tidak diketahui1
Lamidi LamidiAfiliasi tidak diketahui1
Farid AmrinsaniAfiliasi tidak diketahui1
Hafizh Aushaf MahdyAfiliasi tidak diketahui1
Ragimova NazilaAfiliasi tidak diketahui1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://journal.yrpipku.com/index.php/jaets/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Noise Reduction of Motion and EMG Artifacts in Holter ECG Using IIR Filters for Robust Arrhythmia Detection</dc:title>
	<dc:creator xml:lang="en">Sumber, Sumber</dc:creator>
	<dc:creator xml:lang="en">S, Endang Dian</dc:creator>
	<dc:creator xml:lang="en">Triwiyanto, Triwiyanto</dc:creator>
	<dc:creator xml:lang="en">Nashichah, Roichatun</dc:creator>
	<dc:creator xml:lang="en">Athavale, Vijay Anant</dc:creator>
	<dc:subject xml:lang="en">Holter monitor</dc:subject>
	<dc:subject xml:lang="en">IIR Butterworth</dc:subject>
	<dc:subject xml:lang="en">ECG Lead</dc:subject>
	<dc:subject xml:lang="en">Motion Artifact</dc:subject>
	<dc:subject xml:lang="en">Muscle Artifact</dc:subject>
	<dc:description xml:lang="en">Ambulatory Holter electrocardiography (ECG) enables continuous monitoring for detecting transient arrhythmias; however, its diagnostic reliability is significantly degraded by motion artifacts and electromyographic (EMG) interference. Under severe motion artifact conditions, prior studies report that ambulatory ECG SNR can fall below −10 dB , although SNR levels vary substantially depending on activity type and electrode placement, reducing usable data segments and impairing arrhythmia detection. While advanced denoising methods such as wavelet transforms and deep learning achieve high accuracy, their computational complexity limits real-time deployment in resource-constrained embedded systems. This reveals a critical gap in lightweight methods that jointly optimize noise suppression, morphological preservation, and downstream diagnostic performance. This study proposes a computationally efficient IIR Butterworth bandpass filtering framework for real-time IoT-based Holter ECG systems. The system combines three-lead ECG acquisition, embedded processing on an ESP32, and real-time visualization. Performance is assessed using SNR, mean squared error (MSE), Pearson correlation, and confusion matrix-based detection metrics on ten male participants under controlled motion and muscle artifact conditions. Results demonstrate statistically significant SNR improvements for motion artifacts (ΔSNR = 9.47 ± 1.96 dB, t(9) = 15.28, p &amp;lt; 0.001) and EMG artifacts (ΔSNR = 16.73 ± 0.91 dB, t(9) = 58.11, p &amp;lt; 0.0001). Post-filtering morphological fidelity was high, with mean Pearson correlation of 0.963 for motion artifacts and 0.945 for muscle artifacts. These signal quality improvements translated into 95.3% post-filtering arrhythmia detection accuracy (sensitivity: ≈96.0%, specificity: ≥97.0%, F1-score: ≥95.0%), significantly exceeding the 70% minimum performance threshold adopted in this study as a conservative screening criterion (t(9) = 29.7, p &amp;lt; 0.001). Despite dataset limitations (n = 10), the proposed framework provides an effective trade-off between computational efficiency and diagnostic reliability, supporting scalable and real-time ambulatory ECG monitoring for early arrhythmia screening.</dc:description>
	<dc:publisher xml:lang="en">Yayasan Riset dan Pengembangan Intelektual (YRPI)</dc:publisher>
	<dc:date>2026-06-15</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.yrpipku.com/index.php/jaets/article/view/9426</dc:identifier>
	<dc:identifier>10.37385/jaets.v7i2.9426</dc:identifier>
	<dc:source xml:lang="en">Journal of Applied Engineering and Technological Science (JAETS); Vol. 7 No. 2 (2026): Journal of Applied Engineering and Technological Science (JAETS); 1177-1201</dc:source>
	<dc:source>2715-6079</dc:source>
	<dc:source>2715-6087</dc:source>
	<dc:source>10.37385/jaets.v7i2</dc:source>
	<dc:language>en</dc:language>
	<dc:relation>https://journal.yrpipku.com/index.php/jaets/article/view/9426/6394</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2026 Sumber Sumber, Endang Dian S, Triwiyanto Triwiyanto, Roichatun Nashichah, Vijay Anant Athavale</dc:rights>
</oai_dc:dc>