Abstrak

Liquids or medications administered through intravenous systems (infusion) is necessary when a patient requires immediate medication or requires slowly and continuously administered medications. This method is considered adequate in certain situations; however, the liquids/medications administration of through infusion has a risk when the nurse is late in replacing intravenous fluid. In this study, the system was designed to monitor the fluid infusion level by utilizing a short text message system. This system will provide information when the infusion fluid level is at 50 ml, 20 ml, and 0 via SMS. Moreover, a buzzer is utilized that acts as an alarm when the reaching 20 ml level, and infusion fluid have not been replaced yet, and when the intravenous fluid does not drip. Infra-red sensors and photodiodes are used to detect intravenous droplets of fluid, which are then used to calculate fluids' volume. The system is controlled using an Atmega 328 microcontroller. SIM Modem 900 is used to send SMS. Based on tests carried out, infusion fluid level monitoring systems have excellent performance. The percentage of system errors when detecting fluid level infuse is 1,21%. The function of sending fluid level information through SMS also works well

Kata kunci

Jaringan kolaborasi para penulis

7 penulis · 11 tautan kepenulisan bersama · 4 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Muhammadiyah University of Yogyakarta
  • Universiti Malaysia Perlis
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Dewanti Catur DhamayantiMuhammadiyah University of Yogyakarta2
Zulkarnay ZakariaUniversiti Malaysia Perlis1
Yessi JusmanMuhammadiyah University of Yogyakarta1
Siti Nurul Aqmariah Mohd KanafiahUniversiti Malaysia Perlis1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://journal.umy.ac.id/index.php/jrc/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Short Text Message Based Infusion Fluid Level Monitoring System</dc:title>
	<dc:creator>Safitri, Meilia</dc:creator>
	<dc:creator>Da Fonseca, Helena</dc:creator>
	<dc:creator>Loniza, Erika</dc:creator>
	<dc:subject xml:lang="en-US">Infusion fluid level</dc:subject>
	<dc:subject xml:lang="en-US">SIM 900</dc:subject>
	<dc:subject xml:lang="en-US">IR</dc:subject>
	<dc:subject xml:lang="en-US">photodiode</dc:subject>
	<dc:description xml:lang="en-US">Liquids or medications administered through intravenous systems (infusion) is necessary when a patient requires immediate medication or requires slowly and continuously administered medications. This method is considered adequate in certain situations; however, the liquids/medications administration of through infusion has a risk when the nurse is late in replacing intravenous fluid. In this study, the system was designed to monitor the fluid infusion level by utilizing a short text message system. This system will provide information when the infusion fluid level is at 50 ml, 20 ml, and 0 via SMS. Moreover, a buzzer is utilized that acts as an alarm when the reaching 20 ml level, and infusion fluid have not been replaced yet, and when the intravenous fluid does not drip. Infra-red sensors and photodiodes are used to detect intravenous droplets of fluid, which are then used to calculate fluids' volume. The system is controlled using an Atmega 328 microcontroller. SIM Modem 900 is used to send SMS. Based on tests carried out, infusion fluid level monitoring systems have excellent performance. The percentage of system errors when detecting fluid level infuse is 1,21%. The function of sending fluid level information through SMS also works well</dc:description>
	<dc:publisher xml:lang="en-US">Universitas Muhammadiyah Yogyakarta</dc:publisher>
	<dc:date>2021-03-19</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.umy.ac.id/index.php/jrc/article/view/9538</dc:identifier>
	<dc:identifier>10.18196/jrc.2253</dc:identifier>
	<dc:source xml:lang="en-US">Journal of Robotics and Control (JRC); Vol. 2 No. 2 (2021): March; 60-64</dc:source>
	<dc:source>2715-5072</dc:source>
	<dc:source>2715-5056</dc:source>
	<dc:source>10.18196/jrc.v2i2</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://journal.umy.ac.id/index.php/jrc/article/view/9538/5612</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2020 Journal of Robotics and Control (JRC)</dc:rights>
</oai_dc:dc>