Melola Discovery

← Melola Discovery

Jurnal SINTA 1

Design of a Continuous Passive Motion (CPM) Machine for Wrist Joint Therapy

Journal of Robotics and Control (JRC) · Vol. 2 No. 4 (2021) · hlm. 311–315

DOI
10.18196/jrc.2498
ISSN
2715-5072, 2715-5056
Terbit
2021-07-01
Penerbit
Universitas Muhammadiyah Yogyakarta
Bahasa
eng

Abstrak

Patients after joint surgery require passive movements that are carried out continuously. The movement is given to the patient to reduce the stiffness that occurs in the joints after joint surgery. Joint stiffness is characterized by pain and limited motion of the joint, which is caused due to reduced synovial fluid. Synovial fluid in the joints can be reduced if the joint is not moved for a long time. This research will design a wrist joint therapy machine that can be moved flexion, extension, ulnar, and radial. Besides that, the device to be designed can be adjusted the angle of movement of the tool and the speed of movement of the tool. The angles of movement that can be reached are (1) flexion movement: 80 ° movement angle, (2) extension movement: 75 ° movement angle, (3) radial movement: 25 ° movement angle, and (4) ulnar movement: movement angle 39 °. Adjustable speeds include 1 RPM, 2 RPM, and 3 RPM. In the result of the testing device, we find the maximum difference of movement in 2° and the speed of rotation we have the difference in 0.5 seconds. Based on experiments that have been done, the tool can be controlled in accordance with the desired movement settings.

Kata kunci

Jaringan kolaborasi para penulis

3 penulis · 3 tautan kepenulisan bersama

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Sanata Dharma University
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://journal.umy.ac.id/index.php/jrc/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Design of a Continuous Passive Motion (CPM) Machine for Wrist Joint Therapy</dc:title>
	<dc:creator>Noviyanto, Antonius Hendro</dc:creator>
	<dc:creator>Septilianingtyas, Laurentia Deby</dc:creator>
	<dc:creator>Rahmawati, Dita</dc:creator>
	<dc:subject xml:lang="en-US">CPM Machine</dc:subject>
	<dc:subject xml:lang="en-US">Wrist</dc:subject>
	<dc:subject xml:lang="en-US">Joint Stiffness</dc:subject>
	<dc:subject xml:lang="en-US">Rehabilitation</dc:subject>
	<dc:subject xml:lang="en-US">Microcontroller</dc:subject>
	<dc:description xml:lang="en-US">Patients after joint surgery require passive movements that are carried out continuously. The movement is given to the patient to reduce the stiffness that occurs in the joints after joint surgery. Joint stiffness is characterized by pain and limited motion of the joint, which is caused due to reduced synovial fluid. Synovial fluid in the joints can be reduced if the joint is not moved for a long time. This research will design a wrist joint therapy machine that can be moved flexion, extension, ulnar, and radial. Besides that, the device to be designed can be adjusted the angle of movement of the tool and the speed of movement of the tool. The angles of movement that can be reached are (1) flexion movement: 80 ° movement angle, (2) extension movement: 75 ° movement angle, (3) radial movement: 25 ° movement angle, and (4) ulnar movement: movement angle 39 °. Adjustable speeds include 1 RPM, 2 RPM, and 3 RPM. In the result of the testing device, we find the maximum difference of movement in 2° and the speed of rotation we have the difference in 0.5 seconds. Based on experiments that have been done, the tool can be controlled in accordance with the desired movement settings.</dc:description>
	<dc:publisher xml:lang="en-US">Universitas Muhammadiyah Yogyakarta</dc:publisher>
	<dc:date>2021-07-01</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.umy.ac.id/index.php/jrc/article/view/10031</dc:identifier>
	<dc:identifier>10.18196/jrc.2498</dc:identifier>
	<dc:source xml:lang="en-US">Journal of Robotics and Control (JRC); Vol. 2 No. 4 (2021): July; 311-315</dc:source>
	<dc:source>2715-5072</dc:source>