Melola Discovery

← Melola Discovery

Jurnal SINTA 1

Design and Develop a Non-Invasive Pulmonary Vibration Device for Secretion Drainage in Pediatric Patients with Pneumonia

Journal of Robotics and Control (JRC) · Vol. 4 No. 5 (2023) · hlm. 632–642

DOI
10.18196/jrc.v4i5.19588
ISSN
2715-5072, 2715-5056
Terbit
2023-09-14
Penerbit
Universitas Muhammadiyah Yogyakarta
Bahasa
eng

Abstrak

The study aimed to develop a non-invasive pulmonary vibration device, specifically tailored for pediatric patients, to address a range of pulmonary conditions. The device employs a PID control system to ensure consistent and precise vibrations. The primary contribution of this research is the successful development, testing, and implementation of this innovative device. Utilizing technical components such as an Arduino, a vibration DC motor, and an ADXL335 accelerometer, the device was engineered to deliver stable and continuous vibrations even when subjected to external pressures or interactions with the patient. Controllers, including P, PI, PD, and PID types, were rigorously compared. The Ziegler-Nichols tuning technique was applied for meticulous evaluation of vibration control specifically within the context of this non-invasive pulmonary vibration device. Our findings revealed that the PID controller displayed superior accuracy in vibration control compared to P, PI, and PD controllers. Clinical trials involving pediatric patients showed that the PID-controlled device achieved treatment outcomes comparable to conventional methods. The device's precise control of vibration strength provides an added benefit, making it a well-tolerated, non-invasive treatment option for various pulmonary conditions in pediatric patients. Future research is necessary to assess the long-term effectiveness of the device and to facilitate its integration into standard clinical practice. In summary, this study represents a significant advancement in pediatric pulmonary care, demonstrating the critical role that PID control systems adapted for non-invasive pulmonary vibration devices can play in enhancing treatment precision and outcomes.

Kata kunci

Jaringan kolaborasi para penulis

20 penulis · 109 tautan kepenulisan bersama · 33 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Rangsit University
  • Rajamangala University of Technology
  • Mae Fah Luang University
  • Pathumwan Institute of Technology
  • King Mongkut's University of Technology Thonburi
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Tasawan PuttasakulRangsit University18
Pariwat ImuraRangsit University15
Yutthana PititheeraphabRangsit University11
Anuchit NirapaiRangsit University11
Wanida KhotakhamRangsit University9
Pareena DangyaiAfiliasi tidak diketahui7
Ratchanee SaosuwanRangsit University7
Panya MinyongRajamangala University of Technology5
Benjamas PanomruttanarugKing Mongkut's University of Technology Thonburi4
Jaroonrut PrinyakuptRangsit University4
Udomsak JantontapoRajamangala University of Technology3
Songtham DeewanichsakulRajamangala University of Technology3

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://journal.umy.ac.id/index.php/jrc/oai, pertama di-harvest 9 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en-US">Design and Develop a Non-Invasive Pulmonary Vibration Device for Secretion Drainage in Pediatric Patients with Pneumonia</dc:title>
	<dc:creator>Wongkamhang, Anantasak</dc:creator>
	<dc:creator>Wuttipan, Nathamon</dc:creator>
	<dc:creator>Chotikunnan, Rawiphon</dc:creator>
	<dc:creator>Roongprasert, Kittipan</dc:creator>
	<dc:creator>Chotikunnan, Phichitphon</dc:creator>
	<dc:creator>Thongpance, Nuntachai</dc:creator>
	<dc:creator>Sangworasil, Manas</dc:creator>
	<dc:creator>Srisiriwat, Anuchart</dc:creator>
	<dc:subject xml:lang="en-US">Pulmonary</dc:subject>
	<dc:subject xml:lang="en-US">PID Control</dc:subject>
	<dc:subject xml:lang="en-US">Vibration Device</dc:subject>
	<dc:subject xml:lang="en-US">Pneumonia.</dc:subject>
	<dc:description xml:lang="en-US">The study aimed to develop a non-invasive pulmonary vibration device, specifically tailored for pediatric patients, to address a range of pulmonary conditions. The device employs a PID control system to ensure consistent and precise vibrations. The primary contribution of this research is the successful development, testing, and implementation of this innovative device. Utilizing technical components such as an Arduino, a vibration DC motor, and an ADXL335 accelerometer, the device was engineered to deliver stable and continuous vibrations even when subjected to external pressures or interactions with the patient. Controllers, including P, PI, PD, and PID types, were rigorously compared. The Ziegler-Nichols tuning technique was applied for meticulous evaluation of vibration control specifically within the context of this non-invasive pulmonary vibration device. Our findings revealed that the PID controller displayed superior accuracy in vibration control compared to P, PI, and PD controllers. Clinical trials involving pediatric patients showed that the PID-controlled device achieved treatment outcomes comparable to conventional methods. The device's precise control of vibration strength provides an added benefit, making it a well-tolerated, non-invasive treatment option for various pulmonary conditions in pediatric patients. Future research is necessary to assess the long-term effectiveness of the device and to facilitate its integration into standard clinical practice. In summary, this study represents a significant advancement in pediatric pulmonary care, demonstrating the critical role that PID control systems adapted for non-invasive pulmonary vibration devices can play in enhancing treatment precision and outcomes.</dc:description>
	<dc:publisher xml:lang="en-US">Universitas Muhammadiyah Yogyakarta</dc:publisher>
	<dc:date>2023-09-14</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.umy.ac.id/index.php/jrc/article/view/19588</dc:identifier>
	<dc:identifier>10.18196/jrc.v4i5.19588</dc:identifier>
	<dc:source xml:lang="en-US">Journal of Robotics and Control (JRC); Vol. 4 No. 5 (2023); 632-642</dc:source>
	<dc:source>2715-5072</dc:source>
	<dc:source>2715-5056</dc:source>
	<dc:source>10.18196/jrc.v4i5</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://journal.umy.ac.id/index.php/jrc/article/view/19588/8582</dc:relation>
	<dc:relation>https://journal.umy.ac.id/index.php/jrc/article/view/19588/10877</dc:relation>
	<dc:rights xml:lang="en-US">Copyright (c) 2023 Anantasak Wongkamhang, Nathamon Wuttipan, Rawiphon Chotikunnan, Kittipan Roongprasert, Phichitphon Chotikunnan, Nuntachai Thongpance, Manas Sangworasil, Anuchart Srisiriwat</dc:rights>
</oai_dc:dc>