Abstrak

Naphthalene sulfonic acids (NSAs) are widely used as chemical tracers in reservoir characterization due to their stability, low adsorption, and cost-efficiency. This study presents a high‑performance liquid chromatography (HPLC) method with fluorescence detection for the simultaneous separation and quantification of six NSA tracers (2,6‑NDSA, 1,5‑NDSA, 2,7‑NDSA, 1,6‑NDSA, 1,3,5‑NTSA, and 2‑NSA). The optimized method employed a superficially porous C18 column combined with tetrabutylammonium phosphate and phosphate buffer as ion‑pairing agents. Complete chromatographic resolution was achieved with detection limits ranging from 2.05 to 2.97 µg/L and quantification limits between 6.54–8.90 µg/L. Calibration curves exhibited excellent linearity (R2 ≥ 0.999) across 10–1,000 µg/L. Application of this method in an interwell tracer test demonstrated clear tracer breakthrough behavior, enabling differentiation of injector contributions with recovery values ranging from 2 to 67%. These results confirm that the developed method provides a sensitive, selective, and cost‑effective analytical approach for multi‑tracer monitoring under complex reservoir conditions and supports optimization of EOR strategies.

Kata kunci

Jaringan kolaborasi para penulis

13 penulis · 25 tautan kepenulisan bersama

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Pertamina (Indonesia)
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Oktaviani Kusuma WardaniPertamina (Indonesia)4
Dzulhijah Nur MeisincaAfiliasi tidak diketahui1
Rene Indrawan PratamoraPertamina (Indonesia)1
Cece RahayuPertamina (Indonesia)1
Muhammad Daffa RizqullahPertamina (Indonesia)1
ArjunaPertamina (Indonesia)1
Tyas KuswardaniPertamina (Indonesia)1
Erik Indra PutraPertamina (Indonesia)1
Heru SugiartoAfiliasi tidak diketahui1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OpenAlex API (jurnal SINTA, ADR 0048) https://openalex.org/S2764840418, pertama di-harvest 19 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title xml:lang="en">Analytical Method for Multi-Tracer Detection of Naphthalene Sulfonates Using HPLC with Fluorescence Detector for Interwell Injection Pattern Analysis in Enhanced Oil Recovery</dc:title><dc:creator>Ester Tio Minar E. Silalahi</dc:creator><dc:creator>Maulana Hardi</dc:creator><dc:creator>Adi Widiyanto</dc:creator><dc:creator>Muhammad Hatta Nasution</dc:creator><dc:subject>naphthalenesulfonic acids</dc:subject><dc:subject>chemical tracers</dc:subject><dc:subject>interwell tracer test</dc:subject><dc:subject>enhanced oil recovery</dc:subject><dc:subject>fluorescence detection</dc:subject><dc:subject>ion-pair chromatography</dc:subject><dc:subject>tracer breakthrough</dc:subject><dc:subject>reservoir characterization</dc:subject><dc:subject>naphthalene sulfonate</dc:subject><dc:description xml:lang="en">Naphthalene sulfonic acids (NSAs) are widely used as chemical tracers in reservoir characterization due to their stability, low adsorption, and cost-efficiency. This study presents a high‑performance liquid chromatography (HPLC) method with fluorescence detection for the simultaneous separation and quantification of six NSA tracers (2,6‑NDSA, 1,5‑NDSA, 2,7‑NDSA, 1,6‑NDSA, 1,3,5‑NTSA, and 2‑NSA). The optimized method employed a superficially porous C18 column combined with tetrabutylammonium phosphate and phosphate buffer as ion‑pairing agents. Complete chromatographic resolution was achieved with detection limits ranging from 2.05 to 2.97 µg/L and quantification limits between 6.54–8.90 µg/L. Calibration curves exhibited excellent linearity (R2 ≥ 0.999) across 10–1,000 µg/L. Application of this method in an interwell tracer test demonstrated clear tracer breakthrough behavior, enabling differentiation of injector contributions with recovery values ranging from 2 to 67%. These results confirm that the developed method provides a sensitive, selective, and cost‑effective analytical approach for multi‑tracer monitoring under complex reservoir conditions and supports optimization of EOR strategies.</dc:description><dc:publisher>Gadjah Mada University</dc:publisher><dc:date>2026-11-03</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:identifier>https://doi.org/10.22146/ijc.113462</dc:identifier><dc:identifier>10.22146/ijc.113462</dc:identifier><dc:source>Indonesian Journal of Chemistry; Vol 26 No 6 (2026); 1898-1898</dc:source><dc:source>1411-9420</dc:source><dc:source>2460-1578</dc:source><dc:language>en</dc:language><dc:rights>https://creativecommons.org/licenses/by-nc-nd/4.0/</dc:rights></oai_dc:dc>