Correlation between the Methods of Flowcytometry, Manual Light and Fluorescence Microscopes for CD4 Cell Count in Normal Adults

Folia Medica Indonesiana · Vol. 40 No. 3 (2004)

ISSN
2355-8393, 2599-056X
Terbit
2004-07-01
Penerbit
Airlangga University
Bahasa
en

Abstrak

A laboratory study was conducted in order to assess the correlation and to compare between flowcytometry, manual light and fluorescence microscope methods used for enumeration of CD4 cell count. K3EDTA whole blood samples from 15 healthy volunteers were taken for determining CD4 cell count by flowcytometry, manual light and fluorescence microscope methods. The results of the manual light and fluorescence microscope methods showed a significant correlation to the flowcytometry method, r = 0.949 (p = 0.000) and r = 0.959 (p = 0.000) respectively. In addition, the manual light microscope method demonstrated a significant correlation to the manual fluorescence microscope method, r = 0.927 (p = 0.000). The comparison between the flowcytometry method, the manual light and fluorescence microscope methods was not significantly different, respectively p = 0.175 and 0.055. Furthermore, there was no significant difference between the manual light microscope and the fluorescence microscope, p = 0.768. In using the manual light and fluorescence microscope methods, the CD4 cell count were enumerated by three different laboratory technicians. The results showed a significant correlation between the manual light microscope and flowcytometry methods, r = 0.943 (p = 0.000), r = 0.943 (p = 0.000) and r = 0.935 (p = 0.000), respectively. Moreover, compared to the flowcytometry method the results were not significantly different, p = 0.167, 0.121 and 0.412, respectively. The manual fluorescence microscope method showed a significant correlation to the flowcytometry method, r = 0.961 (p = 0.000), r = 0.947 (p = 0.000) and 0.932 (p = 0.000), respectively. The comparison to the flowcytometry method showed that the results of the first and second technician were not significantly different p = 0.642 and 0.052, respectively, however, the result of the third technician was significantly different (p = 0.018). Based on the results of the data analysis obtained in this study, it can be concluded that the manual methods can be

Kata kunci

Jaringan kolaborasi para penulis

37 penulis · 75 tautan kepenulisan bersama · 10 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Universitas Airlangga
  • Chittagong University of Engineering & Technology
  • State University of Malang
  • University of Brawijaya
  • Rumah Sakit Dokter Soetomo
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
SoetojoAfiliasi tidak diketahui2
Erwin Astha TriyonoUniversitas Airlangga2
Windhu PurnomoUniversitas Airlangga2
Ferry EffendiAfiliasi tidak diketahui1
Purwaningsih PurwaningsihUniversitas Airlangga1
Nuzul Qur’aniatiUniversitas Airlangga1
Ganden SupriyantoUniversitas Airlangga1
Hanim Istatik Badi’ahUniversitas Airlangga1
Ni Nyoman Tri PuspaningsihUniversitas Airlangga1
Wahyu Endah PrabawatiUniversitas Airlangga1
Gatut SuhendroUniversitas Airlangga1
Sri Agus SudjarwoUniversitas Airlangga1
Wiwiek Indriyani MaskoepUniversitas Airlangga1
Siti KhairunisaUniversitas Airlangga1
Indro HandojoAfiliasi tidak diketahui1
S P EdijantoAfiliasi tidak diketahui1
Sophia Yudaprawira SalimAfiliasi tidak diketahui1
Saiful IslamChittagong University of Engineering & Technology1
Dwi Lily LukasAfiliasi tidak diketahui1
Muhammad AminuddinRumah Sakit Dokter Soetomo1
Suharjono SuharjonoUniversity of Brawijaya1
Yen Yen Ari IndrawijayaState University of Malang1
Gilang Mauladi RahmanUniversitas Airlangga1
Doddy Moesbadianto SoebadiUniversitas Airlangga1
Hariyono HariyonoUniversitas Airlangga1
Widodo JpAfiliasi tidak diketahui1
Juli SoemarsonoAfiliasi tidak diketahui1
Ety Retno SetyowatiAfiliasi tidak diketahui1
Alberthina SuripattyAfiliasi tidak diketahui1
Hermi Indita MalewaAfiliasi tidak diketahui1
Siswanto DarmadiAfiliasi tidak diketahui1
Magdalena TobingAfiliasi tidak diketahui1
Nurwasis NurwasisUniversitas Airlangga1
Mohamad Nurdin ZuhriUniversitas Airlangga1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OpenAlex API (jurnal SINTA, ADR 0048) https://openalex.org/S4210179714, pertama di-harvest 19 Nov 2026, terakhir terlihat 22 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title xml:lang="en">Correlation between the Methods of Flowcytometry, Manual Light and Fluorescence Microscopes for CD4 Cell Count in Normal Adults</dc:title><dc:creator>Endang Retnowati</dc:creator><dc:creator>Wira Santoso</dc:creator><dc:creator>Nasronudin Nasronudin</dc:creator><dc:subject>CD4 count</dc:subject><dc:subject>flow cytometry</dc:subject><dc:subject>fluorescence microscopy</dc:subject><dc:subject>method comparison</dc:subject><dc:subject>healthy adults</dc:subject><dc:subject>whole blood</dc:subject><dc:subject>HIV diagnostics</dc:subject><dc:subject>assay agreement</dc:subject><dc:subject>method specificity</dc:subject><dc:description xml:lang="en">A laboratory study was conducted in order to assess the correlation and to compare between flowcytometry, manual light and fluorescence microscope methods used for enumeration of CD4 cell count. K3EDTA whole blood samples from 15 healthy volunteers were taken for determining CD4 cell count by flowcytometry, manual light and fluorescence microscope methods. The results of the manual light and fluorescence microscope methods showed a significant correlation to the flowcytometry method, r = 0.949 (p = 0.000) and r = 0.959 (p = 0.000) respectively. In addition, the manual light microscope method demonstrated a significant correlation to the manual fluorescence microscope method, r = 0.927 (p = 0.000). The comparison between the flowcytometry method, the manual light and fluorescence microscope methods was not significantly different, respectively p = 0.175 and 0.055. Furthermore, there was no significant difference between the manual light microscope and the fluorescence microscope, p = 0.768. In using the manual light and fluorescence microscope methods, the CD4 cell count were enumerated by three different laboratory technicians. The results showed a significant correlation between the manual light microscope and flowcytometry methods, r = 0.943 (p = 0.000), r = 0.943 (p = 0.000) and r = 0.935 (p = 0.000), respectively. Moreover, compared to the flowcytometry method the results were not significantly different, p = 0.167, 0.121 and 0.412, respectively. The manual fluorescence microscope method showed a significant correlation to the flowcytometry method, r = 0.961 (p = 0.000), r = 0.947 (p = 0.000) and 0.932 (p = 0.000), respectively. The comparison to the flowcytometry method showed that the results of the first and second technician were not significantly different p = 0.642 and 0.052, respectively, however, the result of the third technician was significantly different (p = 0.018). Based on the results of the data analysis obtained in this study, it can be concluded that the manual methods can be</dc:description><dc:publisher>Airlangga University</dc:publisher><dc:date>2004-07-01</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:identifier>http://journal.unair.ac.id/filerPDF/EndangRE.pdf</dc:identifier><dc:source>Folia Medica Indonesiana; Vol 40 No 3 (2004)</dc:source><dc:source>2355-8393</dc:source><dc:source>2599-056X</dc:source><dc:language>en</dc:language></oai_dc:dc>