Melola Discovery

← Melola Discovery

Open access Jurnal SINTA 1

A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource Settings

International Journal of Electrical and Computer Engineering (IJECE) · Vol. 9 No. 4 (2019)

DOI
10.11591/ijece.v9i4.pp2321-2327
ISSN
2088-8708, 2088-8708
Published
2019-08-01
Publisher
Institute of Advanced Engineering and Science
License
creativecommons.org/licenses/by-nc/4.0

Abstract

Medical devices are often expensive, so people in low-income countries cannot afford them. This paper presents the design of a low-cost wearable medical device to measure vital signs of a patient including heart rate, blood oxygen saturation level (SpO2) and respiratory rate. The wearable medical device mainly consists of a microcontroller and two biomedical sensors including airflow thermal sensor to measure respiratory rate and pulse oximeter sensor to measure SpO2 and heart rate. We can monitor the vital signs from a smartphone using a web browser through IEEE802.11 wireless connectivity to the wearable medical device. Furthermore, the wearable medical device requires simple management to operate; hence, it can be easily used. Performance evaluation results show that the designed wearable medical device works as good as a standard SpO2 device and it can measure the respiratory rate properly. The designed wearable medical device is inexpensive and appropriate for low-resource settings. Moreover, as its components are commonly available in the market, it easy to assembly and repair locally.

Collaboration network of the authors

3 authors · 3 co-authorship links

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Hasanuddin University
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

This record comes from Crossref (what the publisher deposited), because the journal's own OAI-PMH endpoint could not be harvested.

Metadata source: Crossref API (cadangan, berdasarkan ISSN) https://api.crossref.org/journals/2088-8708/works, first harvested 27 Okt 2026, last seen 27 Okt 2026.
Metadata as oai_dc, converted from Crossref
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title>A Low Cost Wearable Medical Device for Vital Signs Monitoring in Low-Resource Settings</dc:title><dc:creator>Niswar, Muhammad</dc:creator><dc:creator>Nur, Muhammad</dc:creator><dc:creator>Mappangara, Idar</dc:creator><dc:description>&lt;jats:p&gt;Medical devices are often expensive, so people in low-income countries cannot afford them. This paper presents the design of a low-cost wearable medical device to measure vital signs of a patient including heart rate, blood oxygen saturation level (SpO2) and respiratory rate. The wearable medical device mainly consists of a microcontroller and two biomedical sensors including airflow thermal sensor to measure respiratory rate and pulse oximeter sensor to measure SpO2 and heart rate. We can monitor the vital signs from a smartphone using a web browser through IEEE802.11 wireless connectivity to the wearable medical device. Furthermore, the wearable medical device requires simple management to operate; hence, it can be easily used. Performance evaluation results show that the designed wearable medical device works as good as a standard SpO2 device and it can measure the respiratory rate properly.  The designed wearable medical device is inexpensive and appropriate for low-resource settings. Moreover, as its components are commonly available in the market, it easy to assembly and repair locally.&lt;/jats:p&gt;</dc:description><dc:publisher>Institute of Advanced Engineering and Science</dc:publisher><dc:date>2019-08-01</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:format>application/pdf</dc:format><dc:identifier>https://doi.org/10.11591/ijece.v9i4.pp2321-2327</dc:identifier><dc:identifier>10.11591/ijece.v9i4.pp2321-2327</dc:identifier><dc:source>International Journal of Electrical and Computer Engineering (IJECE); Vol 9 No 4 (2019); 2321</dc:source><dc:source>2088-8708</dc:source><dc:source>2088-8708</dc:source><dc:relation>http://ijece.iaescore.com/index.php/IJECE/article/viewFile/11180/12836</dc:relation><dc:rights>http://creativecommons.org/licenses/by-nc/4.0</dc:rights></oai_dc:dc>