Melola Discovery

← Melola Discovery

Akses terbuka Jurnal SINTA 1

Denoising MAX6675 reading using Kalman filter and factorial design

International Journal of Electrical and Computer Engineering (IJECE) · Vol. 11 No. 5 (2021)

DOI
10.11591/ijece.v11i5.pp3818-3827
ISSN
2722-2578, 2088-8708
Terbit
2021-10-01
Penerbit
Institute of Advanced Engineering and Science
Lisensi
creativecommons.org/licenses/by-sa/4.0

Abstrak

This paper aims to tune the Kalman filter (KF) input variables, namely measurement error and process noise, based on two-level factorial design. Kalman filter then was applied in inexpensive temperature-acquisition utilizing MAX6675 and K-type thermocouple with Arduino as its microprocessor. Two levels for each input variable, respectively, 0.1 and 0.9, were selected and applied to four K-type thermocouples mounted on MAX6675. Each sensor with a different combination of input variables was used to measure the temperature of ambient-water, boiling water, and sudden temperature drops in the system. The measurement results which consisted of the original and KF readings were evaluated to determine the optimum combination of input variables. It was found that the optimum combination of input variables was highly dependent on the system's dynamics. For systems with relatively constant dynamics, a large value of measurement error and small value of process noise results in higher precision readings. Nevertheless, for fast dynamic systems, the previous input variables' combination is less optimal because it produced a time-gap, which made the KF reading differ from the original measurement. The selection of the optimum input combination using two-level factorial design eased the KF tuning process, resulting in a more precise yet low-cost sensor.

Jaringan kolaborasi para penulis

9 penulis · 17 tautan kepenulisan bersama · 3 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • University of Indonesia
  • Diponegoro University
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
SetyowatiAfiliasi tidak diketahui1
Hanny HandiyaniUniversity of Indonesia1
Diah ArruumUniversity of Indonesia1
Setyowati SetyowatiUniversity of Indonesia1
Heriandi SutadiUniversity of Indonesia1
Anggorowati AnggorowatiDiponegoro University1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Data ini berasal dari Crossref (yang disetorkan penerbit), karena endpoint OAI-PMH milik jurnal tidak dapat di-harvest.

Sumber metadata: Crossref API (cadangan, berdasarkan ISSN) https://api.crossref.org/journals/2088-8708/works, pertama di-harvest 27 Okt 2026, terakhir terlihat 15 Nov 2026.
Metadata sebagai oai_dc, dikonversi dari Crossref
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title>Denoising MAX6675 reading using Kalman filter and factorial design</dc:title><dc:creator>Septiana, Reski</dc:creator><dc:creator>Roihan, Ibnu</dc:creator><dc:creator>Koestoer, Raldi A.</dc:creator><dc:description>&lt;jats:p&gt;This paper aims to tune the Kalman filter (KF) input variables, namely measurement error and process noise, based on two-level factorial design. Kalman filter then was applied in inexpensive temperature-acquisition utilizing MAX6675 and K-type thermocouple with Arduino as its microprocessor. Two levels for each input variable, respectively, 0.1 and 0.9, were selected and applied to four K-type thermocouples mounted on MAX6675. Each sensor with a different combination of input variables was used to measure the temperature of ambient-water, boiling water, and sudden temperature drops in the system. The measurement results which consisted of the original and KF readings were evaluated to determine the optimum combination of input variables. It was found that the optimum combination of input variables was highly dependent on the system's dynamics. For systems with relatively constant dynamics, a large value of measurement error and small value of process noise results in higher precision readings. Nevertheless, for fast dynamic systems, the previous input variables' combination is less optimal because it produced a time-gap, which made the KF reading differ from the original measurement. The selection of the optimum input combination using two-level factorial design eased the KF tuning process, resulting in a more precise yet low-cost sensor.&lt;/jats:p&gt;</dc:description><dc:publisher>Institute of Advanced Engineering and Science</dc:publisher><dc:date>2021-10-01</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:format>application/pdf</dc:format><dc:identifier>https://doi.org/10.11591/ijece.v11i5.pp3818-3827</dc:identifier><dc:identifier>10.11591/ijece.v11i5.pp3818-3827</dc:identifier><dc:source>International Journal of Electrical and Computer Engineering (IJECE); Vol 11 No 5 (2021); 3818</dc:source><dc:source>2722-2578</dc:source><dc:source>2088-8708</dc:source><dc:relation>http://ijece.iaescore.com/index.php/IJECE/article/viewFile/25231/15023</dc:relation><dc:rights>http://creativecommons.org/licenses/by-sa/4.0</dc:rights></oai_dc:dc>