ANALISIS NILAI COMPUTED TOMOGRAPHY DOSE INDEX (CTDI) PADA PHANTOM KEPALA DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH SAYANG RAKYAT MAKASSAR

JOURNAL ONLINE OF PHYSICS · Vol. 10 No. 1 (2024) · hlm. 55–61

DOI
10.22437/jop.v10i1.37149
ISSN
2502-2016
Terbit
2024-11-08
Penerbit
Prodi Fisika FST UNJA
Bahasa
en
Lisensi
creativecommons.org/licenses/by/4.0

Abstrak

This research aims to determine the CTDI value on a CT Scan aircraft using Dose Profiler and to determine the comparison of the CTDI (Computed Tomography Dose Index) value on the surface of each measurement position. The measurement method was carried out by measuring the CT Scan radiation dose with the CTDI value on the phantom using a RaySafe multimeter at a voltage of 130 kV, using a thickness of 10 mm, a time current accuracy of 100 mAs, 200 mAs and 300 mAs. The research results show the dose received on the surface of each type of phantom position with different time current accuracy respectively for a time current of 100 mAs at the center (9.139 mGy), type 1 (9.606 mGy), type 2 (9.170) type 3 (9.221), and type 4 (9,210). A time current of 200 mAs is found at the center (9,110 mGy), edge 1 (9,578 mGy), edge 2 (9,546 mGy), edge 3 (9,224 mGy), and edge 4 (9,200 mGy). A time current of 300 mAs is found at the center position (18.39 mGy), edge 1 (19.33 mGy), edge 2 (19.18 mGy), edge 3 (18.48 mGy), and 4 (18.51 mGy). Time-current variations affect the radiation dose that can be received by the phantom and have a linear influence on the CTDIvol value.

Kata kunci

Jaringan kolaborasi para penulis

7 penulis · 12 tautan kepenulisan bersama · 5 lintas organisasi

Arahkan kursor atau ketuk sebuah nama untuk menyorot rekan penulisnya. Klik dua kali untuk membuka halaman penulis.

  • Universitas Islam Negeri Alauddin Makassar
  • IPB University
  • Universitas Sulawesi Barat
  • Afiliasi tidak diketahui
  • Organisasi berbeda
  • Organisasi sama
  • Afiliasi tidak diketahui
  • Garis lebih tebal = lebih banyak artikel bersama · lingkaran lebih besar = lebih banyak artikel
Tampilkan sebagai tabel
Rekan penulisOrganisasiArtikel bersama
Rahmaniah NalwiUniversitas Sulawesi Barat1
Akhiruddin MadduIPB University1
Mersi KurniatiIPB University1

Afiliasi berasal dari OpenAlex; metadata OAI-PMH jurnal tidak memuatnya. Penulis tanpa ID OpenAlex dicocokkan berdasarkan nama, sehingga dua orang bisa tergabung atau satu orang bisa terpisah.

Sumber metadata: OAI-PMH (metadata jurnal) https://online-journal.unja.ac.id/index.php/jop/oai, pertama di-harvest 6 Nov 2026, terakhir terlihat 16 Nov 2026.
Metadata asli (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">ANALISIS NILAI COMPUTED TOMOGRAPHY DOSE INDEX (CTDI) PADA PHANTOM KEPALA DI INSTALASI RADIOLOGI RUMAH SAKIT UMUM DAERAH SAYANG RAKYAT MAKASSAR</dc:title>
	<dc:creator xml:lang="en">Ilham, Ilham</dc:creator>
	<dc:creator xml:lang="en">Zelviani, Sri</dc:creator>
	<dc:creator xml:lang="en">Jumardin, Jumardin</dc:creator>
	<dc:creator xml:lang="en">Bariah, Khaerul</dc:creator>
	<dc:subject xml:lang="en">CTDI;</dc:subject>
	<dc:subject xml:lang="en">CT Scan;</dc:subject>
	<dc:subject xml:lang="en">Dose Profiler;</dc:subject>
	<dc:subject xml:lang="en">Head Phantom;</dc:subject>
	<dc:description xml:lang="en">This research aims to determine the CTDI value on a CT Scan aircraft using&amp;nbsp; Dose Profiler and to determine the comparison of the CTDI (Computed Tomography Dose Index) value on the surface of each measurement position. The measurement method was carried out by measuring the CT Scan radiation dose with the CTDI value on the phantom using a RaySafe multimeter at a voltage of 130 kV, using a thickness of 10 mm, a time current accuracy of 100 mAs, 200 mAs and 300 mAs. The research results show the dose received on the surface of each type of phantom position with different time current accuracy respectively for a time current of 100 mAs at the center (9.139 mGy), type 1 (9.606 mGy), type 2 (9.170) type 3 (9.221) , and type 4 (9,210). A time current of 200 mAs is found at the center (9,110 mGy), edge 1 (9,578 mGy), edge 2 (9,546 mGy), edge 3 (9,224 mGy), and edge 4 (9,200 mGy). A time current of 300 mAs is found at the center position (18.39 mGy), edge 1 (19.33 mGy), edge 2 (19.18 mGy), edge 3 (18.48 mGy), and 4 (18.51 mGy) . Time-current variations affect the radiation dose that can be received by the phantom and have a linear influence on the CTDIvol value.</dc:description>
	<dc:publisher xml:lang="en">Prodi Fisika FST UNJA</dc:publisher>
	<dc:date>2024-11-08</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://online-journal.unja.ac.id/jop/article/view/37149</dc:identifier>
	<dc:identifier>10.22437/jop.v10i1.37149</dc:identifier>
	<dc:source xml:lang="en">JOURNAL ONLINE OF PHYSICS; Vol. 10 No. 1 (2024): JOP (Journal Online of Physics) Vol 10 No 1; 55-61</dc:source>
	<dc:source>2502-2016</dc:source>
	<dc:language>en</dc:language>
	<dc:relation>https://online-journal.unja.ac.id/jop/article/view/37149/19299</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2024 Ilham Ilham, Sri Zelviani, Jumardin Jumardin, Khaerul Bariah</dc:rights>
	<dc:rights xml:lang="en">https://creativecommons.org/licenses/by/4.0</dc:rights>
</oai_dc:dc>