Are mechanical and electromechanical methods accurately interchangeable for measuring plasma prothrombin time and activated partial thromboplastin time? A comparative analysis study and potential implication to cardiovascular disease

Heart Science Journal · Vol. 6 No. 1 (2025) · pp. 72–76

DOI
10.21776/ub.hsj.2025.006.01.12
ISSN
2721-9984, 2721-9976
Published
2025-01-31
Publisher
Universitas Brawijaya
Language
en
License
creativecommons.org/licenses/by/4.0

Abstract

INTRODUCTION: The DT100 offers both optical and mechanical modes, with its mechanical mode showing better homogenization than the STAGO, but comparative study is limited. OBJECTIVES: The study aimed to evaluate the diagnostic accuracy of plasma Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) measurements using the DT100 and STAGO instruments. METHODS: Designated as a cross-sectional study, this study was conducted at RSUD Dr. Soetomo from October 2022 to January 2023. Venous blood samples with plasma citrate anticoagulant 0.109 M 3.2% were consecutively collected from hospitalized patients, and all samples underwent testing using both the DT100 in mechanical mode (DT100, TCoag Ireland Limited, Ireland) and the STAGO employing an electromechanical method (Compact Max3, STAGO, France). Statistical analysis included comparison using Paired t-test, Pearson correlation, and Bland-Altman analysis to assess agreement between the results obtained from the two instruments. RESULTS: The study included 51 patients. PT levels were significantly lower with the DT100 compared to STAGO (MD: -2.0; 95%CI: (-2.30) – (-1.3); p<0.0001), and showed a strong positive correlation between methods (r:0.9535; p<0.0001). However, Bland-Altman analysis for PT showed a bias of 1.84, with limits of agreement (3.30-0.37), indicating systematic differences and variability. APTT levels were significantly higher with DT100 compared to STAGO (MD:3.60; 95%CI: 2.13–5.07; p<0.0001), with a moderate positive correlation (r:0.6690; p<0.0001). For APTT, bias of Bland-Altman analysis was -3.60, with limits ((-9.84) – (2.64)), suggesting significant discrepancies and variability between methods. CONCLUSION: The study found significant variability in PT and APTT measurements between the DT100 and STAGO methods.

Keywords

Collaboration network of the authors

15 authors · 42 co-authorship links · 8 across different organisations

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Universitas Airlangga
  • University of Surabaya
  • Affiliation unknown
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
Jusak NugrahaUniversitas Airlangga2
Abdul Hafid BajamalUniversitas Airlangga2
Valentinus BesinUniversity of Surabaya2
Mohammad Hasan MachfoedUniversitas Airlangga2
Nur Aisiyah WidjajaUniversitas Airlangga1
Gwyneth Trixie StanpoUniversitas Airlangga1
Langgeng Agung WaskitoUniversitas Airlangga1
Pepy Dwi EndraswariUniversitas Airlangga1
Mangestuti AgilUniversitas Airlangga1
I Ketut SudianaUniversitas Airlangga1
Naesilla NaesillaUniversitas Airlangga1
Budi UtomoUniversitas Airlangga1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

Metadata source: OAI-PMH (metadata jurnal) https://heartscience.ub.ac.id/index.php/heartscience/oai, first harvested 8 Nov 2026, last seen 16 Nov 2026.
Original metadata (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Are mechanical and electromechanical methods accurately interchangeable for measuring plasma prothrombin time and activated partial thromboplastin time? A comparative analysis study and potential implication to cardiovascular disease</dc:title>
	<dc:creator xml:lang="en">Pranata, Fransiska Jaya</dc:creator>
	<dc:creator xml:lang="en">Notopuro, Paulus Budiono</dc:creator>
	<dc:creator xml:lang="en">Hajat, Arifoel</dc:creator>
	<dc:subject xml:lang="en">Prothrombin time</dc:subject>
	<dc:subject xml:lang="en">activated partial thromboplastin time</dc:subject>
	<dc:subject xml:lang="en">mechanical method</dc:subject>
	<dc:subject xml:lang="en">DT100 TCoag</dc:subject>
	<dc:subject xml:lang="en">STAGO Compact Max 3</dc:subject>
	<dc:description xml:lang="en">INTRODUCTION: The DT100 offers both optical and mechanical modes, with its mechanical mode showing better homogenization than the STAGO, but comparative study is limited.
OBJECTIVES: The study aimed to evaluate the diagnostic accuracy of plasma Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) measurements using the DT100 and STAGO instruments.
METHODS: Designated as a cross-sectional study, this study was conducted at RSUD Dr. Soetomo from October 2022 to January 2023. Venous blood samples with plasma citrate anticoagulant 0.109 M 3.2% were consecutively collected from hospitalized patients, and all samples underwent testing using both the DT100 in mechanical mode (DT100, TCoag Ireland Limited, Ireland) and the STAGO employing an electromechanical method (Compact Max3, STAGO, France). Statistical analysis included comparison using Paired t-test, Pearson correlation, and Bland-Altman analysis to assess agreement between the results obtained from the two instruments.
RESULTS: The study included 51 patients. PT levels were significantly lower with the DT100 compared to STAGO (MD: -2.0; 95%CI: (-2.30) – (-1.3); p&amp;lt;0.0001), and showed a strong positive correlation between methods (r:0.9535; p&amp;lt;0.0001). However, Bland-Altman analysis for PT showed a bias of 1.84, with limits of agreement (3.30-0.37), indicating systematic differences and variability. APTT levels were significantly higher with DT100 compared to STAGO (MD:3.60; 95%CI: 2.13–5.07; p&amp;lt;0.0001), with a moderate positive correlation (r:0.6690; p&amp;lt;0.0001). For APTT, bias of Bland-Altman analysis was -3.60, with limits ((-9.84) – (2.64)), suggesting significant discrepancies and variability between methods.
CONCLUSION: The study found significant variability in PT and APTT measurements between the DT100 and STAGO methods.</dc:description>
	<dc:publisher xml:lang="en">Universitas Brawijaya</dc:publisher>
	<dc:date>2025-01-31</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://heartscience.ub.ac.id/index.php/heartscience/article/view/20186</dc:identifier>
	<dc:identifier>10.21776/ub.hsj.2025.006.01.12</dc:identifier>
	<dc:source xml:lang="en">Heart Science Journal; Vol. 6 No. 1 (2025): Challenges in Managing Acute Heart Failure; 72-76</dc:source>
	<dc:source>2721-9984</dc:source>
	<dc:source>2721-9976</dc:source>
	<dc:source>10.21776/ub.hsj.2025.006.01</dc:source>
	<dc:language>en</dc:language>
	<dc:relation>https://heartscience.ub.ac.id/index.php/heartscience/article/view/20186/472</dc:relation>
	<dc:relation>10.21776/</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2025 Fransiska Jaya Pranata, Paulus Budiono Notopuro, Arifoel Hajat</dc:rights>
	<dc:rights xml:lang="en">http://creativecommons.org/licenses/by/4.0</dc:rights>
</oai_dc:dc>