Melola Discovery

← Melola Discovery

Journal SINTA 1

Dosimetric Analysis of 6 MV Energy Photon Radiation Beam on Flatness and Symmetry on Linac in Radiotherapy Installation of UNAND Hospital

JURNAL ILMIAH PENDIDIKAN FISIKA AL BIRUNI · Vol. 13 No. 2 (2024) · pp. 175–189

DOI
10.24042/jipfalbiruni.v13i2.24016
ISSN
2503-023X, 2303-1832
Published
2024-12-30
Publisher
Universitas Islam Negeri Raden Intan Lampung, Indonesia
Language
eng

Abstract

A dosimetric analysis of a 6 MV photon energy radiation beam on flatness and symmetry in LINAC radiotherapy at UNAND Hospital has been conducted. The aim was to measure the quality of the photon beam using an ion chamber cc 13 detector by observing the beam profile and analyzing the effect of field size and irradiation depth. This study used a blue phantom as the object of irradiation with 6 MV photon energy and variations in irradiation field size. The results showed that the PDD curve at 6 MV energy was by the international standard recommended by BJR-25. Then, the average values of flatness and symmetry in the field area of 10×10 cm2 and 15×15 cm2 are 1.6% and 2.4% for flatness and 2.3% and 1.9% for symmetry. Thus, the dose distribution is more uniform, and both the left and right sides of the profile on the center axis appear balanced, which will help deliver the dose to the patient better. Thus, these values are suitable for clinical use. This study found that the beam profile was larger on the left side of the main axis. This must be considered to ensure the dose distribution complies with the established safety standards. The results also show that variations in the field area and irradiation depth can affect the beam profile, and the resulting flatness and symmetry values are in accordance with the IAEA TRUS-381 and AAPM TG-142 recommendation standards, which are ±2% flatness and ±3% symmetry.

Keywords

Collaboration network of the authors

15 authors · 37 co-authorship links · 16 across different organisations

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Andalas University
  • Siemens Healthcare (United States)
  • Universitas Gadjah Mada
  • Badan Pengawas Tenaga Nuklir
  • Bani-Waleed University
  • Rumah Sakit Umum Pusat Dr. M. Djamil Padang
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
Dian MilvitaAndalas University3
Mawanda AlmuhayarAndalas University2
Almahdi MousaBani-Waleed University2
Latifah Aulia RasyadaAndalas University2
Kusminarto KusminartoUniversitas Gadjah Mada1
Rani DelvihardiniAndalas University1
Ida Bagus Gede Putra PratamaBadan Pengawas Tenaga Nuklir1
Amel OktaviaAndalas University1
Atika MaulidaAndalas University1
Dina Arfiani RusjdiRumah Sakit Umum Pusat Dr. M. Djamil Padang1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

Metadata source: OAI-PMH (metadata jurnal) http://ejournal.radenintan.ac.id/index.php/al-biruni/oai, first harvested 9 Nov 2026, last seen 22 Nov 2026.
Original metadata (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">Dosimetric Analysis of 6 MV Energy Photon Radiation Beam on Flatness and Symmetry on Linac in Radiotherapy Installation of UNAND Hospital</dc:title>
	<dc:creator xml:lang="en">Analia, Rika</dc:creator>
	<dc:creator xml:lang="en">Fardela, Ramacos</dc:creator>
	<dc:creator xml:lang="en">Ridwan, Ridwan</dc:creator>
	<dc:creator xml:lang="en">Diyona, Fiqi</dc:creator>
	<dc:creator xml:lang="en">Muttaqin, Afdhal</dc:creator>
	<dc:subject xml:lang="en">LINAC</dc:subject>
	<dc:subject xml:lang="en">Dosimetry</dc:subject>
	<dc:subject xml:lang="en">Photon Beam</dc:subject>
	<dc:subject xml:lang="en">Flatness and Symmetry.</dc:subject>
	<dc:description xml:lang="en">A dosimetric analysis of a 6 MV photon energy radiation beam on flatness and symmetry in LINAC radiotherapy at UNAND Hospital has been conducted. The aim was to measure the quality of the photon beam using an ion chamber cc 13 detector by observing the beam profile and analyzing the effect of field size and irradiation depth. This study used a blue phantom as the object of irradiation with 6 MV photon energy and variations in irradiation field size.  The results showed that the PDD curve at 6 MV energy was by the international standard recommended by BJR-25. Then, the average values of flatness and symmetry in the field area of 10×10 cm2 and 15×15 cm2 are 1.6% and 2.4% for flatness and 2.3% and 1.9% for symmetry. Thus, the dose distribution is more uniform, and both the left and right sides of the profile on the center axis appear balanced, which will help deliver the dose to the patient better. Thus, these values are suitable for clinical use. This study found that the beam profile was larger on the left side of the main axis. This must be considered to ensure the dose distribution complies with the established safety standards. The results also show that variations in the field area and irradiation depth can affect the beam profile, and the resulting flatness and symmetry values are in accordance with the IAEA TRUS-381 and AAPM TG-142 recommendation standards, which are ±2% flatness and ±3% symmetry.</dc:description>
	<dc:publisher xml:lang="en">Universitas Islam Negeri Raden Intan Lampung, Indonesia</dc:publisher>
	<dc:date>2024-12-30</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en">Peer-reviewed Article</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/24016</dc:identifier>
	<dc:identifier>10.24042/jipfalbiruni.v13i2.24016</dc:identifier>
	<dc:source xml:lang="en">Jurnal Ilmiah Pendidikan Fisika Al-Biruni; Vol. 13 No. 2 (2024): Jurnal Ilmiah Pendidikan Fisika Al-Biruni; 175-189</dc:source>
	<dc:source xml:lang="id">Jurnal Ilmiah Pendidikan Fisika Al-Biruni; Vol 13 No 2 (2024): Jurnal Ilmiah Pendidikan Fisika Al-Biruni; 175-189</dc:source>
	<dc:source>2503-023X</dc:source>
	<dc:source>2303-1832</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://ejournal.radenintan.ac.id/index.php/al-biruni/article/view/24016/pdf</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2024 Physics Education Department, UIN Raden Intan Lampung, Indonesia</dc:rights>
</oai_dc:dc>