Correlation between the Methods of Flowcytometry, Manual Light and Fluorescence Microscopes for CD4 Cell Count in Normal Adults

Folia Medica Indonesiana · Vol. 40 No. 3 (2004)

ISSN
2355-8393, 2599-056X
Published
2004-07-01
Publisher
Airlangga University
Language
en

Abstract

A laboratory study was conducted in order to assess the correlation and to compare between flowcytometry, manual light and fluorescence microscope methods used for enumeration of CD4 cell count. K3EDTA whole blood samples from 15 healthy volunteers were taken for determining CD4 cell count by flowcytometry, manual light and fluorescence microscope methods. The results of the manual light and fluorescence microscope methods showed a significant correlation to the flowcytometry method, r = 0.949 (p = 0.000) and r = 0.959 (p = 0.000) respectively. In addition, the manual light microscope method demonstrated a significant correlation to the manual fluorescence microscope method, r = 0.927 (p = 0.000). The comparison between the flowcytometry method, the manual light and fluorescence microscope methods was not significantly different, respectively p = 0.175 and 0.055. Furthermore, there was no significant difference between the manual light microscope and the fluorescence microscope, p = 0.768. In using the manual light and fluorescence microscope methods, the CD4 cell count were enumerated by three different laboratory technicians. The results showed a significant correlation between the manual light microscope and flowcytometry methods, r = 0.943 (p = 0.000), r = 0.943 (p = 0.000) and r = 0.935 (p = 0.000), respectively. Moreover, compared to the flowcytometry method the results were not significantly different, p = 0.167, 0.121 and 0.412, respectively. The manual fluorescence microscope method showed a significant correlation to the flowcytometry method, r = 0.961 (p = 0.000), r = 0.947 (p = 0.000) and 0.932 (p = 0.000), respectively. The comparison to the flowcytometry method showed that the results of the first and second technician were not significantly different p = 0.642 and 0.052, respectively, however, the result of the third technician was significantly different (p = 0.018). Based on the results of the data analysis obtained in this study, it can be concluded that the manual methods can be

Keywords

Collaboration network of the authors

37 authors · 75 co-authorship links · 10 across different organisations

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Universitas Airlangga
  • Chittagong University of Engineering & Technology
  • State University of Malang
  • University of Brawijaya
  • Rumah Sakit Dokter Soetomo
  • Affiliation unknown
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
SoetojoAffiliation unknown2
Erwin Astha TriyonoUniversitas Airlangga2
Windhu PurnomoUniversitas Airlangga2
Ferry EffendiAffiliation unknown1
Purwaningsih PurwaningsihUniversitas Airlangga1
Nuzul Qur’aniatiUniversitas Airlangga1
Ganden SupriyantoUniversitas Airlangga1
Hanim Istatik Badi’ahUniversitas Airlangga1
Ni Nyoman Tri PuspaningsihUniversitas Airlangga1
Wahyu Endah PrabawatiUniversitas Airlangga1
Gatut SuhendroUniversitas Airlangga1
Sri Agus SudjarwoUniversitas Airlangga1
Wiwiek Indriyani MaskoepUniversitas Airlangga1
Siti KhairunisaUniversitas Airlangga1
Indro HandojoAffiliation unknown1
S P EdijantoAffiliation unknown1
Sophia Yudaprawira SalimAffiliation unknown1
Saiful IslamChittagong University of Engineering & Technology1
Dwi Lily LukasAffiliation unknown1
Muhammad AminuddinRumah Sakit Dokter Soetomo1
Suharjono SuharjonoUniversity of Brawijaya1
Yen Yen Ari IndrawijayaState University of Malang1
Gilang Mauladi RahmanUniversitas Airlangga1
Doddy Moesbadianto SoebadiUniversitas Airlangga1
Hariyono HariyonoUniversitas Airlangga1
Widodo JpAffiliation unknown1
Juli SoemarsonoAffiliation unknown1
Ety Retno SetyowatiAffiliation unknown1
Alberthina SuripattyAffiliation unknown1
Hermi Indita MalewaAffiliation unknown1
Siswanto DarmadiAffiliation unknown1
Magdalena TobingAffiliation unknown1
Nurwasis NurwasisUniversitas Airlangga1
Mohamad Nurdin ZuhriUniversitas Airlangga1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

Metadata source: OpenAlex API (jurnal SINTA, ADR 0048) https://openalex.org/S4210179714, first harvested 19 Nov 2026, last seen 22 Nov 2026.
Original metadata (oai_dc)
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title xml:lang="en">Correlation between the Methods of Flowcytometry, Manual Light and Fluorescence Microscopes for CD4 Cell Count in Normal Adults</dc:title><dc:creator>Endang Retnowati</dc:creator><dc:creator>Wira Santoso</dc:creator><dc:creator>Nasronudin Nasronudin</dc:creator><dc:subject>CD4 count</dc:subject><dc:subject>flow cytometry</dc:subject><dc:subject>fluorescence microscopy</dc:subject><dc:subject>method comparison</dc:subject><dc:subject>healthy adults</dc:subject><dc:subject>whole blood</dc:subject><dc:subject>HIV diagnostics</dc:subject><dc:subject>assay agreement</dc:subject><dc:subject>method specificity</dc:subject><dc:description xml:lang="en">A laboratory study was conducted in order to assess the correlation and to compare between flowcytometry, manual light and fluorescence microscope methods used for enumeration of CD4 cell count. K3EDTA whole blood samples from 15 healthy volunteers were taken for determining CD4 cell count by flowcytometry, manual light and fluorescence microscope methods. The results of the manual light and fluorescence microscope methods showed a significant correlation to the flowcytometry method, r = 0.949 (p = 0.000) and r = 0.959 (p = 0.000) respectively. In addition, the manual light microscope method demonstrated a significant correlation to the manual fluorescence microscope method, r = 0.927 (p = 0.000). The comparison between the flowcytometry method, the manual light and fluorescence microscope methods was not significantly different, respectively p = 0.175 and 0.055. Furthermore, there was no significant difference between the manual light microscope and the fluorescence microscope, p = 0.768. In using the manual light and fluorescence microscope methods, the CD4 cell count were enumerated by three different laboratory technicians. The results showed a significant correlation between the manual light microscope and flowcytometry methods, r = 0.943 (p = 0.000), r = 0.943 (p = 0.000) and r = 0.935 (p = 0.000), respectively. Moreover, compared to the flowcytometry method the results were not significantly different, p = 0.167, 0.121 and 0.412, respectively. The manual fluorescence microscope method showed a significant correlation to the flowcytometry method, r = 0.961 (p = 0.000), r = 0.947 (p = 0.000) and 0.932 (p = 0.000), respectively. The comparison to the flowcytometry method showed that the results of the first and second technician were not significantly different p = 0.642 and 0.052, respectively, however, the result of the third technician was significantly different (p = 0.018). Based on the results of the data analysis obtained in this study, it can be concluded that the manual methods can be</dc:description><dc:publisher>Airlangga University</dc:publisher><dc:date>2004-07-01</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:identifier>http://journal.unair.ac.id/filerPDF/EndangRE.pdf</dc:identifier><dc:source>Folia Medica Indonesiana; Vol 40 No 3 (2004)</dc:source><dc:source>2355-8393</dc:source><dc:source>2599-056X</dc:source><dc:language>en</dc:language></oai_dc:dc>