Hemoglobin A1c (HbA1c) Levels from the Examination of Different Blood Proportions in K2EDTA Tubes using an Enzymatic Method

Folia Medica Indonesiana · Vol. 59 No. 3 (2023) · pp. 262–266

DOI
10.20473/fmi.v59i3.40652
ISSN
2355-8393, 2599-056X
Published
2023-09-10
Publisher
Airlangga University
Language
en
License
creativecommons.org/licenses/by-nc-sa/4.0/

Abstract

Highlights: 1. Research on the effect of blood volume proportion in the examination of HbA1c levels using K2EDTA anticoagulantvhas yet to be widely carried out in Indonesia.2. The analysis conducted utilizing K2EDTA tubes revealed that the varying proportions of blood samples had no effectvon HbA1c levels.3. This article highlights the necessity of evaluating the pre-analytical phase (sample preparation) of a laboratory process to improve the accuracy of the results and minimize false high or low results in the HbA1c test. Abstract Hemoglobin A1c (HbA1c) examination is the gold standard for diagnosing and monitoring diabetes mellitus patients. In the examination, the pre-analytical phase has the most considerable error rate at 61%. One of the contributing factors is the use of anticoagulants that do not adhere to established guidelines. Additionally, the incompatibility of the sample volume proportions and the anticoagulants in K2EDTA tubes has been observed in numerous cases. The significance of HbA1c testing, particularly in the prevention of diabetes complications, underscores the need for appropriate procedures to be followed throughout the sampling and pre-analytical phases. This study aimed to determine the effect of differences in the proportion of blood sample volume and anticoagulants in K2EDTA tubes on HbA1c levels. This research was conducted in August–September 2022. The research samples were collected from the staff and students at the Department of Medical Laboratory Technology, Politeknik Kesehatan Kemenkes Surabaya, Surabaya, Indonesia. The blood samples were divided into K2EDTA tubes with varying volumes of 1 mL, 2 mL, 3 mL, and 4 mL. The HbA1c levels were then examined at the Clinical Chemistry Laboratory of Politeknik Kesehatan Kemenkes Surabaya. The data were analyzed using a one-way ANOVA test. The statistical test results (p>0.05) indicated that the proportion of samples containing anticoagulants in the K2EDTA tubes did not have any significant effect on HbA1c levels. In conclusion, it is acceptable to utilize K2EDTA tubes with varying blood sample volumes for the measurement of HbA1c levels.

Keywords

Collaboration network of the authors

5 authors · 6 co-authorship links · 2 across different organisations

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Poltekkes Kemenkes Surabaya
  • Kementerian Kesehatan Republik Indonesia
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
Evy Diah WoelansariPoltekkes Kemenkes Surabaya1
Dwi KrihariyaniPoltekkes Kemenkes Surabaya1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

Metadata source: OpenAlex API (jurnal SINTA, ADR 0048) https://openalex.org/S4210179714, first harvested 19 Nov 2026, last seen 22 Nov 2026.
Original metadata (oai_dc)
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title xml:lang="en">Hemoglobin A1c (HbA1c) Levels from the Examination of Different Blood Proportions in K2EDTA Tubes using an Enzymatic Method</dc:title><dc:creator>Museyaroh Museyaroh</dc:creator><dc:creator>Musholli Himmatun Nabilah</dc:creator><dc:creator>Retno Werdiningsih</dc:creator><dc:subject>HbA1c</dc:subject><dc:subject>blood sample volume</dc:subject><dc:subject>anticoagulants</dc:subject><dc:subject>pre-analytical phase</dc:subject><dc:subject>sample preparation</dc:subject><dc:subject>diabetes mellitus</dc:subject><dc:subject>enzyme activity assay</dc:subject><dc:subject>laboratory quality control</dc:subject><dc:description xml:lang="en">Highlights: 1. Research on the effect of blood volume proportion in the examination of HbA1c levels using K2EDTA anticoagulantvhas yet to be widely carried out in Indonesia.2. The analysis conducted utilizing K2EDTA tubes revealed that the varying proportions of blood samples had no effectvon HbA1c levels.3. This article highlights the necessity of evaluating the pre-analytical phase (sample preparation) of a laboratory process to improve the accuracy of the results and minimize false high or low results in the HbA1c test. Abstract Hemoglobin A1c (HbA1c) examination is the gold standard for diagnosing and monitoring diabetes mellitus patients. In the examination, the pre-analytical phase has the most considerable error rate at 61%. One of the contributing factors is the use of anticoagulants that do not adhere to established guidelines. Additionally, the incompatibility of the sample volume proportions and the anticoagulants in K2EDTA tubes has been observed in numerous cases. The significance of HbA1c testing, particularly in the prevention of diabetes complications, underscores the need for appropriate procedures to be followed throughout the sampling and pre-analytical phases. This study aimed to determine the effect of differences in the proportion of blood sample volume and anticoagulants in K2EDTA tubes on HbA1c levels. This research was conducted in August–September 2022. The research samples were collected from the staff and students at the Department of Medical Laboratory Technology, Politeknik Kesehatan Kemenkes Surabaya, Surabaya, Indonesia. The blood samples were divided into K2EDTA tubes with varying volumes of 1 mL, 2 mL, 3 mL, and 4 mL. The HbA1c levels were then examined at the Clinical Chemistry Laboratory of Politeknik Kesehatan Kemenkes Surabaya. The data were analyzed using a one-way ANOVA test. The statistical test results (p&gt;0.05) indicated that the proportion of samples containing anticoagulants in the K2EDTA tubes did not have any significant effect on HbA1c levels. In conclusion, it is acceptable to utilize K2EDTA tubes with varying blood sample volumes for the measurement of HbA1c levels.</dc:description><dc:publisher>Airlangga University</dc:publisher><dc:date>2023-09-10</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:format>application/pdf</dc:format><dc:identifier>https://doi.org/10.20473/fmi.v59i3.40652</dc:identifier><dc:identifier>10.20473/fmi.v59i3.40652</dc:identifier><dc:source>Folia Medica Indonesiana; Vol 59 No 3 (2023); 262-266</dc:source><dc:source>2355-8393</dc:source><dc:source>2599-056X</dc:source><dc:language>en</dc:language><dc:relation>https://e-journal.unair.ac.id/FMI/article/download/40652/27184</dc:relation><dc:rights>https://creativecommons.org/licenses/by-nc-sa/4.0/</dc:rights></oai_dc:dc>