FINITE ELEMENT STUDY ON ANGULATED ABUTMENTS IN MAXILLARY POSTERIOR DENTAL IMPLANTS

Jurnal Pendidikan Teknik Mesin · Vol. 26 No. 01 (2026) · pp. 22–31

DOI
10.15294/jptm.v26i01.29582
ISSN
2503-1759, 2503-1759
Published
2026-08-10
Publisher
Universitas Negeri Semarang
Language
eng

Abstract

Dental implants in the posterior maxilla often require angulated abutments due to anatomical limitations. However, increased abutment angulation may influence stress distribution and compromise mechanical integrity. This study aimed to evaluate the biomechanical behavior of implant components with 15°, 17°, and 20° abutment angulations using three-dimensional finite element analysis (FEA). A static vertical load of 276 N was applied to a D4-type bone block model. Results showed that higher abutment angles led to increased von Mises stress and strain, particularly on the abutment U-profile and the screw head. Despite the rising stress, all configurations maintained safety factors above 1.5. These findings indicate that angled abutments up to 20° are mechanically safe under axial loading conditions, supporting their use in clinically challenging posterior maxillary regions.

Keywords

Collaboration network of the authors

40 authors · 164 co-authorship links · showing the 40 most connected

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • State University of Semarang
  • Affiliation unknown
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
Febri Budi DarsonoAffiliation unknown9
JamariAffiliation unknown8
Khoirul HudaAffiliation unknown7
DhiaulhaqueAffiliation unknown6
Ruben Bayu KristiawanAffiliation unknown5
Septian Eko CahyantoAffiliation unknown5
Mohd Syahmi JamaludinAffiliation unknown4
Rachael AndikaAffiliation unknown4
Andrean Rachman RizaldyAffiliation unknown4
Beni Aji SubektiAffiliation unknown4
Muhammad Iqbal RamadhaniAffiliation unknown4
Muhammad KhumaediAffiliation unknown3
Wirawan SumbodoAffiliation unknown3
Heri YudionoAffiliation unknown2
gunagung syahrul nizam sulkanAffiliation unknown2
Fatkhu AmanullohAffiliation unknown2
Dzikra Adi PratamaAffiliation unknown2
Dennis Andhara PutraAffiliation unknown2
Nurul FatullohAffiliation unknown2
Alifian Bagas SaputraAffiliation unknown2
Rizki SetiadiAffiliation unknown2
Khoirudin FathoniState University of Semarang1
AbdurrahmanAffiliation unknown1
MUHAMMAD NUR FAIZINAffiliation unknown1
Atta Putra HarjantoAffiliation unknown1
R. Imanu Danar HerunandiAffiliation unknown1
R. Ambar Kuntoro Mursit GendroyonoAffiliation unknown1
Nugroho Fajar SusantoAffiliation unknown1
Muhammad Wifaqurrokhman AbdissalamAffiliation unknown1
Ari Dwi Nur Indriawan MusyonoAffiliation unknown1
Muhamad Reza Wahyu PrasetyoAffiliation unknown1
Tegar Oki PrasdikaAffiliation unknown1
Imam Fitrah HadiAffiliation unknown1
Prastiyo prasAffiliation unknown1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

Metadata source: OAI-PMH (metadata jurnal) https://journal.unnes.ac.id/journals/JPTM/oai, first harvested 5 Nov 2026, last seen 5 Nov 2026.
Original metadata (oai_dc)
<oai_dc:dc xmlns="http://www.openarchives.org/OAI/2.0/" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xsi:schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/  http://www.openarchives.org/OAI/2.0/oai_dc.xsd">
	<dc:title xml:lang="en">FINITE ELEMENT STUDY ON ANGULATED ABUTMENTS IN MAXILLARY POSTERIOR DENTAL IMPLANTS</dc:title>
	<dc:creator xml:lang="en">Fadhil Putra Ramadhan</dc:creator>
	<dc:creator xml:lang="en">Kriswanto</dc:creator>
	<dc:creator xml:lang="en">Kaleb Priyanto</dc:creator>
	<dc:creator xml:lang="en">Rahmat Doni Widodo</dc:creator>
	<dc:creator xml:lang="en">J. Jamari</dc:creator>
	<dc:creator xml:lang="en">Athanasius Priharyoto Bayuseno</dc:creator>
	<dc:subject xml:lang="en">Abutment angulation</dc:subject>
	<dc:subject xml:lang="en">Dental implant</dc:subject>
	<dc:subject xml:lang="en">Finite element analysis</dc:subject>
	<dc:subject xml:lang="en">Posterior maxilla</dc:subject>
	<dc:subject xml:lang="en">Von Mises stress</dc:subject>
	<dc:description xml:lang="en">Dental implants in the posterior maxilla often require angulated abutments due to anatomical limitations. However, increased abutment angulation may influence stress distribution and compromise mechanical integrity. This study aimed to evaluate the biomechanical behavior of implant components with 15°, 17°, and 20° abutment angulations using three-dimensional finite element analysis (FEA). A static vertical load of 276 N was applied to a D4-type bone block model. Results showed that higher abutment angles led to increased von Mises stress and strain, particularly on the abutment U-profile and the screw head. Despite the rising stress, all configurations maintained safety factors above 1.5. These findings indicate that angled abutments up to 20° are mechanically safe under axial loading conditions, supporting their use in clinically challenging posterior maxillary regions.</dc:description>
	<dc:publisher xml:lang="en">Universitas Negeri Semarang</dc:publisher>
	<dc:date>2026-08-10</dc:date>
	<dc:type>info:eu-repo/semantics/article</dc:type>
	<dc:type>info:eu-repo/semantics/publishedVersion</dc:type>
	<dc:type xml:lang="en">Peer-reviewed Article</dc:type>
	<dc:format>application/pdf</dc:format>
	<dc:identifier>https://journal.unnes.ac.id/journals/JPTM/article/view/29582</dc:identifier>
	<dc:identifier>10.15294/jptm.v26i01.29582</dc:identifier>
	<dc:source xml:lang="en">Jurnal Pendidikan Teknik Mesin; Vol. 26 No. 01 (2026): June 2026 "Special issues for finite element analysis"; 22 - 31</dc:source>
	<dc:source>2503-1759</dc:source>
	<dc:source>2503-1759</dc:source>
	<dc:source>10.15294/jptm.v26i01</dc:source>
	<dc:language>eng</dc:language>
	<dc:relation>https://journal.unnes.ac.id/journals/JPTM/article/view/29582/11029</dc:relation>
	<dc:rights xml:lang="en">Copyright (c) 2026 Jurnal Pendidikan Teknik Mesin</dc:rights>
</oai_dc:dc>