Abstract

Data collection using thermocouple sensors in low-cost data acquisition is prone to noise interference, which could reduce the data quality. Noise sources such as cold junction compensators, electromagnetic interference, and Johnson noise can significantly affect the reliability and accuracy of conventional measurements. This study aims to improve the quality of thermocouple sensor readings on low-cost data acquisition using calibration method based on deep learning and the denoising process using a wavelet transform. This taken approach successfully increase the accuracy value of 97.67% with a mean absolute error (MAE) of 0.2. The precision also increases of 262.7% as indicated by the result of signal-to-noise ratio (SNR) with a value of 105.29 dB. Comparative analysis was carried out against National Instruments® device and it was found that deep-wavelet method had a lower and higher of MAE and SNRdB values of 16.67% and 0.8% respectively. This study shows that the denoising-calibration method with deep-wavelet can improve the accuracy and reliability of data from low-cost thermocouple devices.

Collaboration network of the authors

13 authors · 30 co-authorship links · 4 across different organisations

Hover or tap a name to highlight its co-authors. Double-click to open the author's page.

  • Universitas Pembangunan Nasional Veteran Jakarta
  • Politeknik Negeri Bali
  • Kementerian Energi dan Sumber Daya Mineral
  • Muhammadiyah University of Surakarta
  • Affiliation unknown
  • Different organisations
  • Same organisation
  • Affiliation unknown
  • Thicker line = more joint articles · bigger circle = more articles
Show as table
Co-authorOrganisationJoint articles
Cahyo Setyo WibowoKementerian Energi dan Sumber Daya Mineral1
Imansyah Ibnu HakimAffiliation unknown1
Bambang SuryawanAffiliation unknown1
I Made Kartika DAffiliation unknown1
I Made RastaAffiliation unknown1
I Gusti Ketut PujaAffiliation unknown1
Ngafwan NgafwanMuhammadiyah University of Surakarta1
I Nyoman SuamirAffiliation unknown1

Affiliations come from OpenAlex; journal OAI-PMH metadata carries none. Authors without an OpenAlex ID are matched by name, which can merge or split people.

This record comes from Crossref (what the publisher deposited), because the journal's own OAI-PMH endpoint could not be harvested.

Metadata source: Crossref API (cadangan, berdasarkan ISSN) https://api.crossref.org/journals/2088-8708/works, first harvested 27 Oct 2026, last seen 15 Nov 2026.
Metadata as oai_dc, converted from Crossref
<oai_dc:dc xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:oai_dc="http://www.openarchives.org/OAI/2.0/oai_dc/"><dc:title>Enhancing the accuracy of low-cost thermocouple devices through deep-wavelet neural network calibration</dc:title><dc:creator>Julian, James</dc:creator><dc:creator>Wahyuni, Fitri</dc:creator><dc:creator>Dewantara, Annastya Bagas</dc:creator><dc:creator>Winarta, Adi</dc:creator><dc:creator>Putra, Nandy</dc:creator><dc:description>&lt;jats:p&gt;Data collection using thermocouple sensors in low-cost data acquisition is prone to noise interference, which could reduce the data quality. Noise sources such as cold junction compensators, electromagnetic interference, and Johnson noise can significantly affect the reliability and accuracy of conventional measurements. This study aims to improve the quality of thermocouple sensor readings on low-cost data acquisition using calibration method based on deep learning and the denoising process using a wavelet transform. This taken approach successfully increase the accuracy value of 97.67% with a mean absolute error (MAE) of 0.2. The precision also increases of 262.7% as indicated by the result of signal-to-noise ratio (SNR) with a value of 105.29 dB. Comparative analysis was carried out against National Instruments® device and it was found that deep-wavelet method had a lower and higher of MAE and SNRdB values of 16.67% and 0.8% respectively. This study shows that the denoising-calibration method with deep-wavelet can improve the accuracy and reliability of data from low-cost thermocouple devices.&lt;/jats:p&gt;</dc:description><dc:publisher>Institute of Advanced Engineering and Science</dc:publisher><dc:date>2024-06-01</dc:date><dc:type>info:eu-repo/semantics/article</dc:type><dc:format>application/pdf</dc:format><dc:identifier>https://doi.org/10.11591/ijece.v14i3.pp2625-2633</dc:identifier><dc:identifier>10.11591/ijece.v14i3.pp2625-2633</dc:identifier><dc:source>International Journal of Electrical and Computer Engineering (IJECE); Vol 14 No 3 (2024); 2625</dc:source><dc:source>2722-2578</dc:source><dc:source>2088-8708</dc:source><dc:relation>https://ijece.iaescore.com/index.php/IJECE/article/viewFile/33936/17405</dc:relation><dc:rights>http://creativecommons.org/licenses/by-sa/4.0</dc:rights></oai_dc:dc>